Remove the vial from the sample chamber and
empty the vial.
Add
one DPD No. 1 tablet
straight from the foil and crush the tablet
using a clean stirring rod.
Transfer the contents of the first vial (Glycine solution) into the pre-
pared vail.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the marks
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Chlorine dioxide.
Messtoleranzen:
0 – 1,9 mg/l: ± 0,1 mg/l
> 1,9 – 3,8 mg/l: ± 0,2 mg/l
> 3,8 – 5,7 mg/l: ± 0,4 mg/l
> 5,7 – 7,6 mg/l: ± 0,6 mg/l
> 7,6 – 11 mg/l: ± 0,8 mg/l
RESULT
Zero
Test
CLO
Notes:
1. For the differentiated determination of Chlorine and Chlorine dioxide see method: Chlo-
rine with Tablet, section d, Chlorine with Chlorine dioxide
2. see page 38