3/4
Calibration
Free chlorine (Cl2) calibration
The free chlorine calibration is always a two-point calibration
process: zero point followed by DPD calibration.
During the calibration procedure, feeder output signals can be
maintained at constant output by enabling the "Hold function" first.
Zero-point calibration
1
Press the symbol.
2
Press the menu Calibration.
3
Tap the measurement “Chlorine”.
4
Press the parameter “Zero point”.
5
Close the ball valve on the sample water inlet.
Make sure that the chlorine sensor is firmly screwed in. Otherwi-
se the measurement accuracy will be affected by inhomoge-
neous flow and inadequate sand cleaning. When the sample
water supply has been stopped, the display first drops rapidly,
and after approximately one minute slowly approaches zero.
During first commissioning, it is essential to wait for 5 minutes,
even if the display shows “0.00” or flashes after a few seconds.
6
Wait until the displayed chlorine value no longer changes.
7
Press the “Calibration” button. An input field opens.
8
Press "Enter" to save the zero point.
9
Open the ball valve on the sample water inlet.
Measuring value calibration (DPD)
10
After zero point calibration, wait at least 2 minutes.
11
Open the flow cell drain (drain) by approximately 1 turn and
extract a specimen of the sample water.
12
Determine the content of free chlorine in the sample using
a photometer.
13
Press the parameter “DPD”.
14
Press the “Calibration” button. An input field opens.
15
Use the input keys to enter the determined value.
16
Press "Enter" to save the entry.
pH calibration
During pH calibration, the buffer solution and the sample water
should have the same temperature. If there is a temperature dif-
ference of > 8 °C, first bring the buffer solution to the same tem-
perature as the pool water.
1
Press the symbol.
2
Press the menu Calibration.
3
Tap the measurement “pH”.
pH 7 calibration
4
Press the parameter “pH 7”.
5
Close the sample water inlet and sample water outlet and
briefly open the flow cell drain (drain) to release the pres-
sure. Close the flow cell drain (drain) again.
6
Place the beaker into the bottom clip and fill the beaker with
buffer solution "pH 7.00" or use the bag with buffer solution
"pH 7.00".
7
Unscrew the pH sensor from the flow cell.
8
Dip the pH sensor through the top clip at least 2 cm deep
into the buffer solution and move slightly until the indicated
pH value remains constant.
9
Press the “Calibration” button. An input field opens.
10
Use the input field to enter the value to be calibrated for the
buffer solution.
11
Press "Enter" to save the entry.
pH X-span calibration
12
Remove the sensor from the buffer solution “pH 7.00”.
13
Wash the sensor in distilled water to prevent carryover of
buffer solution.
14
Press the parameter “pH X”.
15
Place the beaker into the bottom clip and fill the beaker with
buffer solution "pH 4.65" or use the bag with buffer solution
"pH 4.65".
If buffer solutions other than those stated are used, the pH value
of the buffer solution must be lower than pH 6 or higher than pH 8.
16
Dip the pH sensor at least 2 cm deep into the buffer solution
and move gently until the indicated pH value remains cons-
tant.
17
Press the “Calibration” button. An input field opens.
18
Use the keypad to enter the value to be calibrated.
19
Press "Enter" to save the entry.
20
Remove the pH sensor from the top clip.
21
Screw the pH sensor into the cover of the cell body of the
flow cell.
22
Open the sample water inlet and outlet again.
Offset compensation
If external influences result in a constant difference between
the displayed pH value and a pH value measured manually,
this difference can be compensated and the comparative value
entered in the Offset menu.
1
Press the symbol.
2
Press the menu Calibration.
3
Tap the measurement “pH”.
4
Press the parameter “Offset”.
5
Use the keypad to enter the value from the comparative
measurement.
6
Press "Enter" to save the entry.
The offset entry is deleted each time a new pH-7 calibration or
span calibration is performed.