Evo Aeroneb Go Instruction Manual Download Page 7

ASSEMBLY

1. Connect the Medication Cup 

(containing the OnQ

Aerosol 

Generator) to the Nebulizer Body.

NOTE:

Make sure the Aeroneb

®

Go is unplugged before you add

medication to the Medication Cup.

2. Fill the Medication Cup with the 

prescribed nebulizer medication.
Never pour more than 6 mL into the 
cup. The continuous line on the inside 
surface of the cup indicates the 
maximum filling level. (When the 
medication cup contains medication,
the cup should remain in an upright 
position).

CAUTION:

To avoid damage to the nebulizer unit, do not use a

syringe with needle.

4. Place the Cap on the Medication Cup.

5. Attach Mouthpiece or Mask Connector

and mask to Nebulizer Body.

NOTE:

Masks are an optional accessory and may be purchased 

through 

evo

Medical Solutions or your Aeroneb

®

Go dealer.

13

6

VINEGAR SOLUTION: (Cont.)

III. Soak the Nebulizer Unit parts (Medication Cup and Cap,

Nebulizer Body, and Mouthpiece or Mask Adapter and Mask) 
in the solution for thirty (30) minutes.

IV. Rinse all pieces thoroughly under running hot tap water.

V. Shake off excess water, and allow parts to air dry on a clean,

dry towel, out of the reach of children.

VI. Make certain that all disinfected parts are completely dry 

before you store them.

CONTROL lll GERMICIDE:

If you would like to disinfect your nebulizer, you may use
the medical disinfectant described below, available from your
supplier.

Use only Control lll Laboratory Germicide (Maril Products Inc.) or
an equivalent medical disinfectant to disinfect your nebulizer.

l. First, clean the Nebulizer Unit as described in the Daily 

Cleaning Section, following Steps 1-3.

Il. Disinfect the Nebulizer Unit parts (Medication Cup and Cap,

Nebulizer Body, and Mouthpiece or Mask Connector Elbow 
and Mask) according to the manufacturer instructions for the 
medical disinfectant.

lll. Clean the Nebulizer Unit again as described in the Daily 

Cleaning Section to remove any residual Control lll.

lV. Shake off excess water, and allow parts to air dry on a clean,

dry towel, out of the reach of children.

V. Make certain that all disinfected parts are completely dry 

before you store them.

Summary of Contents for Aeroneb Go

Page 1: ...ed or tampered with during the warranty period Shipping charges if any shall be paid by the purchaser NOTE There is no other express warranty Implied warranties including those of merchantability and fitness for a particular purpose are limited to the duration of the express limited warranty and to the extent permitted by law and all implied warranties are excluded This warranty does not cover pro...

Page 2: ...tion Manual Thoroughly read and understand the information contained in these instructions before using the Aeroneb Go For future reference SAVE AND REFER TO THESE INSTRUCTIONS REPLACEMENT PARTS AND ACCESSORIES PART REORDER NUMBER 7000 Battery Controller with Cable 21105 7030 Wall Mount Adapter 21099 7070 Battery Controller with Cable 21105 Wall Mount Adapter 21099 7090 Battery Control Module 2221...

Page 3: ...is manual Never operate this product if any of the parts are not working properly or have been damaged Do not use the Aeroneb Go Nebulizer in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide Do not use the Aeroneb Go Nebulizer with anesthetic breathing systems or lung ventilator breathing systems 17 2 PERFORMANCE Aerosol flow rate Greater than 0 25 mL per minute Resi...

Page 4: ...eat a life threatening condition a backup device is necessary ELECTROMAGNETIC SUSCEPTIBILITY This device meets the requirements of IEC 60601 1 2 which describes requirements for electromagnetic compatibility EMC of medical devices This includes immunity to radio frequency electric fields and electrostatic discharge in addition to the other applicable requirements of the standard Compliance with EM...

Page 5: ...f the low battery power light is flashing Replace the batteries or use the AC Powered Controller Check that the Battery Powered Controller cable is connected properly to the nebulizer and to the Control Module Connect the cable properly to the nebulizer and the Battery Powered Controller Refer to page 7 No visible aerosol when using AC Powered Controller Check that the AC Powered Controller is plu...

Page 6: ...ct is being used to treat a life threatening condition a backup device is necessary In event of loss of electrical power remove the AC Powered Controller cable from the Aeroneb Go Nebulizer Unit and operate with the Battery Powered Controller refer to page 7 Never clean your Nebulizer Unit with solvents chemicals alcohol or abrasive cleaners Do not microwave any parts Do not use brushes sharp tool...

Page 7: ...hirty 30 minutes IV Rinse all pieces thoroughly under running hot tap water V Shake off excess water and allow parts to air dry on a clean dry towel out of the reach of children VI Make certain that all disinfected parts are completely dry before you store them CONTROL lll GERMICIDE If you would like to disinfect your nebulizer you may use the medical disinfectant described below available from yo...

Page 8: ... and Mouthpiece or Mask Adapter and Mask by hand in a solution of warm water and mild liquid dishwashing soap Use soap intended only for washing dishes by hand Do not use dishwasher detergent as it is too abrasive Do not leave parts soaking in soapy water for more than 20 minutes 2 Rinse all pieces thoroughly under running hot tap water Make sure that no soap is left on any of the parts 3 Shake of...

Page 9: ...end of the cable after each use Clean the Nebulizer Unit and associated parts daily NOTE Removal and improper reinsertion of the retaining ring from the underside of the medication cup may cause the unit to leak or malfunction 11 8 CLEANING AND STORING YOUR AERONEB GO TO OPTIMIZE THE PERFORMANCE and life expectancy of your Aeroneb Go device we recommend you follow these cleaning instructions caref...

Page 10: ...O OPERATE WITH AN AC DC POWERED CONTROLLER The nebulizer can be powered directly with an AC Powered Controller NOTE Aerosol mask is not provided 1 Insert the AC Powered Controller cable connector into the back of the Nebulizer Unit by sliding the cable straight up the nebulizer body 2 Plug the AC Powered Controller into a standard AC electrical outlet using the appropriate outlet adapter The green...