E2608-VOC
Rev 05.03.2021
Sensor probe handling
The wall-mount version of the detector-transmitter is available with a remote probe (see
drawing below for dimensions). The remote probe is connected to the main unit with a
shielded cable. The default remote probe cable length is 3 m.
A)
Wall mount remote probe with fixing clamp (default version),
B)
Remote probe with rubber flange and three self-tapping screws (on request)
The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to
protect the sensor from dust, dirt, and water drops. The filter should be replaced if it gets
strongly contaminated. To replace the PTFE filter, unscrew the sensor head cap and
remove the old filter. Place a new filter into the cap and tighten it again
.
NOTE! Never stab or press the filter near its center where the sensor is located since this
may damage the sensor. Do not remove the filter as it may cause the device to show
incorrect values and/or break the sensor.
The recommended orientation of the sensor probe is vertical with the sensor tip pointing
downwards. This prevents the possible accumulation of condensed water on the sensor
protection filter.
NOTE! Metal oxide semiconductor sensors are not replaceable.
15