E2608-PID
Rev 05.03.2021
Re-fitting the PID electrode stack and lamp
NOTE! Never refit a damaged lamp.
1.
Place the lamp inside the O-ring seal in the pellet as illustrated below. Twisting the
lamp slightly during insertion will help to ensure the lamp window is snug against
the electrode stack’s front electrode. The lamp should be freely supported by the
O-ring.
2.
Continuing to hold the electrode stack between forefinger and thumb, carefully
insert the lamp into the recess in the sensor ensuring that the lamp remains in
position. Press the electrode stack firmly, to ensure that the electrode stack wing
clips are engaged, and the top faces of the electrode stack and sensor house are
flush.
3.
Refit the sensor into the sensing equipment.
4.
Recalibrate the gas detector.
Discarding the PID electrode stack
The electrode stack does not have any toxic components, however, if it has been
contaminated by toxic materials, show due care when disposing of.