Specifications
Detected gases
VOCs with ionisation potential < 10.6 eV
Sampling method
Diffusion
Sensor type
Photoionization detector
Calibration
Isobutylene
Order code
E2618-PID-40
E2618-PID-200
Typical detection ranges
0...40 ppm (isobutylene)
0...200 ppm (isobutylene)
Resolution / digital unit
1 ppb
1 ppm
Response time T90
< 3 s
Sensor lifetime
5 years
Calibration interval
Monthly or more frequently depending on operating conditions
Signal update
Every 1 second
Load resistance
R
L
< (Us - 2 V) / 22 mA for 4-20 mA
R
L
> 250 kOhm for 0-10 V mode
Digital interface
RS485, Modbus RTU protocol
No galvanic isolation
Power supply
11...30 VDC or 24 VAC
Power consumption
< 2 VA
Analog outputs
2 × 4-20 mA / 0-10 V, user settable
Outputs assignment
OUT1 2 gas; OUT2 2 gas
Cable connections
Screwless spring-loaded terminals
Enclosure
Grey ABS, wall mount, protection class IP65
Dimensions
H82 × W80 × D55 mm
CE marking
According to 2014/30/EU and 2014/35/EU, EN 50491-4-1:2012
EN 61000-6-3:2020, EN 61326-1:2013(EMC, emissions)
EN 61000-6-1:2019, EN 61000-6-2:2019(EMC, Immunity)
EN 60079-29-1:2016, EN 60079-29-2:2015 and EN 60079-29-3:2014
Operating conditions
-40...+65ºC; 0,9...1,1 atm
0...95% RH non-condensing,
Explosion-safe indoor areas, Non-aggressive atmosphere
Other options
Remote probe
Protection IP65, default cable length 3.0 m;
max height 80 mm, max diameter 65 mm
10