![Evikon E2230 User Manual Download Page 8](http://html1.mh-extra.com/html/evikon/e2230/e2230_user-manual_2446023008.webp)
E2230
Rev 09.03.2021
Operation
Powering on the device
After powering on the device following 3 procedures shall take place on the device:
1.
For 3 seconds the device will switch on all LED indicators to make sure of their
functionality.
2.
E2230 will measure one time temperature, humidity and CO2 concentration and
they will be indicated on the panel to make sure sensors work. These values shall
be shown until the device shall be put into indication/logging mode. Logger 2
seconds LEDs shall blink independently to indicate that logger is in
configuration/checking mode.
3.
E2230 will read out logger internal memory and shall confirm memory validation.
This process can take up to 10 seconds depending on how much data has been
saved.
After these procedures logger shall start indicating and logging if datalogging is permitted.
(Depending if circular memory overwriting is enabled or not) E2230 series 2 seconds LEDs
shall blink together after all procedures are finished to indicate indication/logging mode.
Configuration and Communication
1. via WiFi
After powering on the device, it will switch on its own WiFi Hotspot, called E2830-XXXX.
XXXX - Last 4 digits of the Serial No.
Example of WiFi Hotspot name: E2830-1086
Default password of Hotspot WiFi network: E2830E2830
Device will keep up the hotspot for 5 minutes after which it will switch off the hotspot.
8