13
12
“Ramp Up” and “Ramp Down” time.
Therefore, the setting of ON Time should be
no less than two times of the “Ramp” time in
this mode.
ON TIME
≥
Ramp up + Ramp down
20 Alternate Mode The stimulation of the CH2 will occur after
the 1st contraction of CH1 is completed. In
this mode, the setting of ON Time should be
no less than two times of the “Ramp” time.
The OFF Time should be equal or more than
the ON Time.
ON TIME
≥
Ramp up + Ramp down
OFF TIME
≥
ON TIME
21 Timer
Adjustable, from 1 to 60 minutes or
Continuous. Adjustable in 1 minute each
step from 1 to 15 minutes, and 5 minutes
each step from 15 to 60 minutes. Treatment
time countdown automatically.
22 Patient
This unit can store 60 sets of operation
Compliance Meter records. Total recorded time is 999 hours.
23 Low Battery
A low battery indicator will show up when
Indicator
the battery is low.
24 Operating
Temperature: 0°~40
℃
Condition
Relative Humidity:30%~75%
Atmosphere Pressure : 700hPa~1060hPa
.
25 Remark
There may be up to a +/-5% tolerance of
all parameters and +/-20% tolerance of
output amplitude & voltage.
The waveforms of the TENS modes are as follows.
1. Burst
2. Normal
3. Modulation
4. SD1 (Strength-Duration)
5. SD2 (Strength-Duration)