Chapter 18: APPLICATION OF RE-USABLE SELF

ADHESIVE ELECTRODES

Application

1.  Clean and dry the skin at the prescribed area thoroughly with

soap and water prior to application of electrodes.

2. Insert the lead wire into the pin connector on the pre-wired

electrodes.

3. Remove the electrodes from the protective liner and apply the

electrodes firmly to the treatment site.  Make sure that the unit is

      turned off prior to applying the electrodes.

Removal

1.  Turn off the unit prior to removing the electrodes.
2. Lift at the edge of electrodes and peel; do not pull on the lead

wires because it may damage the electrodes.

3. Place the electrodes on the liner and remove the lead wire by

twisting and pulling at the same time.

Care  and  Storage

1. Between uses, store the electrodes in the resealable bag in a
     cool dry place.
2. It may be helpful to improve repeated application by spreading a

few drops of cold water over the adhesive and turn the surface
up to air dry. Over Saturation with water will reduce the adhe-
sive properties.

Chapter 17:  TIPS FOR SKIN CARE

To avoid skin irritation, especially if you have sensitive skin, follow
these suggestions:

1. Wash the area of skin where you will be placing the electrodes,

using mild soap and water before applying electrodes, and after
taking them off.  Be sure to rinse soap off thoroughly and dry skin
well.

2. Excess hair may be clipped with scissors; do not shave stimula-

tion area.

3. Wipe the area with the skin preparation your clinician has

recommended.  Let this dry.  Apply electrodes as directed.

4. Many skin problems arise from the 

pulling stress

 from adhesive

patches that are excessively stretched across the skin during
application.  To prevent this, apply electrodes from center outward;
avoid stretching over the skin.

5. To minimize 

pulling stress

, tape extra lengths of lead wires to

the skin in a loop to prevent tugging on electrodes.

6. When removing electrodes, always remove by pulling in the di-

rection of hair growth.

7. It may be helpful to rub skin lotion on electrode placement area

when not wearing   electrodes.

8. Never apply electrodes over irritated or broken skin.

Summary of Contents for Comfy EMS

Page 1: ...0434 Comfy EMS R FOR THE INSTRUCTION MANUAL INSTRUCTION MANUAL Distributed by V1 1...

Page 2: ...chment of Electrodes Lead Wires 12 14 Lead Wire Maintenance 13 15 Electrode Options 13 16 Electrode Placement 13 17 Tips For Skin Care 14 18 Application of Re usable Self Adhesive Electrodes 15 19 Adj...

Page 3: ...l Muscle Stimulation is an internationally accepted and proven way of treating muscular injuries It works by sending electronic pulses to the muscle needing treatment this causes the muscle to exercis...

Page 4: ...should be used only with the leads and electrodes recommended for use by the manufacturer 10 Portable powered muscle stimulators should not be used while driving operating machinery or during any act...

Page 5: ...ASE 11 BELTCLIP SIDE 12 PROTECTIVE COVER BACK SIDE FRONT 1 LIQUIDCRYSTICALDISPLAY 2 PANELCOVER 3 LEADCONNECTOR 4 INTENSITY CONTROL ON OFF SWITCH 5 MODECONTROL 6 SET CONTROL 7 INCREMENT CONTROL 8 DECRE...

Page 6: ...chnical specification details of Comfy EMS are as follows MECHANISM TECHICALDESCRIPTION 01 Channel Dual isolated between channels 02 PulseAmplitude Adjustable 0 100 mA into 500 ohm load each channel 0...

Page 7: ...erminals in the device Make sure that both Intensity control ON OFF Switch knobs are in the off position 2 Insert the lead wires into the lead wire sockets on top of the device 3 Open the electrode pa...

Page 8: ...conventional settings suggested here If the initial results are not positive speak to your physician about alternative stimulation settings and or electrode placements Once an acceptable placement has...

Page 9: ...dry Over Saturation with water will reduce the adhe sive properties Chapter 17 TIPS FOR SKIN CARE To avoid skin irritation especially if you have sensitive skin follow these suggestions 1 Wash the are...

Page 10: ...lectrodes are intended for single patient use only 4 If irritation occurs discontinue use and consult your clinician 5 Read the instructions for use of self adhesive electrodes before application Chap...

Page 11: ...er When pressing this button the parameter will decrease 8 Timer The unit has a timer of 1 60 minutes and Continue It can be adjusted by pressing the Set and Increment or Decrement controls The treatm...

Page 12: ...ert the battery into the compartment 5 Note the polarity indicated on the battery and in the compartment 6 Replace the battery compartment cover and slide to close f Set Pulse Width The pulse width de...

Page 13: ...e should be stored and transported under the temperature range of 20 C 60 C relative humidity 20 5 Aatmosphere pressure 500hPa 1060hPa Chapter 20 BATTERY INFORMATION PRECATIONS 1 Remove battery if equ...

Page 14: ...tion when necessary Chapter 22 SAFETY TECHNICAL CONTROLS For safety reasons check your Comfy EMS each week based on the following checklist 1 Check the device for external damage deformation of the ho...

Page 15: ...l EMS 1 Constant Mode Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Scope A VERT 10 0V DIV HORIZ 100 s OUTPUT 42 2V pk pk Pulse Rate 148Hz Scope B VERT 10 0V DIV HORIZ 100 s OUTPUT 42 4950Vpk pk Pul...

Page 16: ...00ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relation Time 12 Sec Ramp Time 6 Sec 2 Synchronous Mode Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relatio...

Page 17: ......

Reviews: