background image

17

16

3. Lead Connector

Connection of the electrodes is made with the two lead wires.

     The device must be switched off before connecting the cables.
     Both intensity controls must be at the Off -position.  Electrodes
      must be pressed firmly on the skin.

4. Mode Control

There are 3 modes available 

 Constant, Synchronous and

Alternate.  A mode can be selected by pressing the 

Mode

 control.

5. Set Control

By pressing the 

SET

 control, you may enter the setting you

intend to make adjustment. You may start to set the value by
pressing the 

Increment

 and 

Decrement

 controls when the

value is flashing.

6. Increment Control

This button controls the increase of settings.  When pressing this
button, the parameter will increase.

Important

1. Do not apply to broken skin.
2. The electrodes should be discarded when they are no longer

adhering.

3. The electrodes are intended for single patient use only.
4. If irritation occurs, discontinue use and consult your clinician.
5. Read the instructions for use of self-adhesive electrodes before

application.

Chapter 19:  ADJUSTING THE CONTROLS

1. Panel Cover:

A lid covers the controls for selecting mode and adjusting settings.
Your medical professional may wish to set these controls for you
and request that you leave the cover in place.

2. Power On/Off Switch and Intensity Controls:

If both controls are in the off-position, the device is switched off.
By turning the controls clockwise, the appropriate channel is
switched on and the indicator of power (CH1 or CH2) will reveal
on the LCD.

The current strength of the impulses transmitted to the electrodes
increases further when the control is turned clockwise.
To reduce the current strength or switch the device off, turn the
control counter clockwise to the required setting or off-position,
respectively.The controls are protected by a cap to avoid

      unintentional change of intensity.

Summary of Contents for Comfy EMS

Page 1: ...0434 Comfy EMS R FOR THE INSTRUCTION MANUAL INSTRUCTION MANUAL Distributed by V1 1...

Page 2: ...chment of Electrodes Lead Wires 12 14 Lead Wire Maintenance 13 15 Electrode Options 13 16 Electrode Placement 13 17 Tips For Skin Care 14 18 Application of Re usable Self Adhesive Electrodes 15 19 Adj...

Page 3: ...l Muscle Stimulation is an internationally accepted and proven way of treating muscular injuries It works by sending electronic pulses to the muscle needing treatment this causes the muscle to exercis...

Page 4: ...should be used only with the leads and electrodes recommended for use by the manufacturer 10 Portable powered muscle stimulators should not be used while driving operating machinery or during any act...

Page 5: ...ASE 11 BELTCLIP SIDE 12 PROTECTIVE COVER BACK SIDE FRONT 1 LIQUIDCRYSTICALDISPLAY 2 PANELCOVER 3 LEADCONNECTOR 4 INTENSITY CONTROL ON OFF SWITCH 5 MODECONTROL 6 SET CONTROL 7 INCREMENT CONTROL 8 DECRE...

Page 6: ...chnical specification details of Comfy EMS are as follows MECHANISM TECHICALDESCRIPTION 01 Channel Dual isolated between channels 02 PulseAmplitude Adjustable 0 100 mA into 500 ohm load each channel 0...

Page 7: ...erminals in the device Make sure that both Intensity control ON OFF Switch knobs are in the off position 2 Insert the lead wires into the lead wire sockets on top of the device 3 Open the electrode pa...

Page 8: ...conventional settings suggested here If the initial results are not positive speak to your physician about alternative stimulation settings and or electrode placements Once an acceptable placement has...

Page 9: ...dry Over Saturation with water will reduce the adhe sive properties Chapter 17 TIPS FOR SKIN CARE To avoid skin irritation especially if you have sensitive skin follow these suggestions 1 Wash the are...

Page 10: ...lectrodes are intended for single patient use only 4 If irritation occurs discontinue use and consult your clinician 5 Read the instructions for use of self adhesive electrodes before application Chap...

Page 11: ...er When pressing this button the parameter will decrease 8 Timer The unit has a timer of 1 60 minutes and Continue It can be adjusted by pressing the Set and Increment or Decrement controls The treatm...

Page 12: ...ert the battery into the compartment 5 Note the polarity indicated on the battery and in the compartment 6 Replace the battery compartment cover and slide to close f Set Pulse Width The pulse width de...

Page 13: ...e should be stored and transported under the temperature range of 20 C 60 C relative humidity 20 5 Aatmosphere pressure 500hPa 1060hPa Chapter 20 BATTERY INFORMATION PRECATIONS 1 Remove battery if equ...

Page 14: ...tion when necessary Chapter 22 SAFETY TECHNICAL CONTROLS For safety reasons check your Comfy EMS each week based on the following checklist 1 Check the device for external damage deformation of the ho...

Page 15: ...l EMS 1 Constant Mode Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Scope A VERT 10 0V DIV HORIZ 100 s OUTPUT 42 2V pk pk Pulse Rate 148Hz Scope B VERT 10 0V DIV HORIZ 100 s OUTPUT 42 4950Vpk pk Pul...

Page 16: ...00ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relation Time 12 Sec Ramp Time 6 Sec 2 Synchronous Mode Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relatio...

Page 17: ......

Reviews: