TROUBLE SHOOTING GUIDE
Problem
Cause
Solution
Press ‘ON’ key but no
display
a) Batteries not in place
b) Batteries not in correct polarity
(+ and – position).
c) Weak
batteries
a) Check that batteries are in place
and making good contact.
b) Re-insert batteries with correct
polarity.
c) Replace
batteries.
Not responding to key
press
a) HOLD mode in operation.
b) Damaged
key
pad.
c) Internal
program
error.
a)
Cancel HOLD mode by
pressing Hold key.
b)
Return to dealer.
c)
Reset all internal programs by
reinserting batteries.
Unstable readings
a)
Insufficient electrolyte in probe.
b)
Air bubbles in probe.
c) Dirty
probe.
d)
Probe not deep enough in
sample.
e)
External noise pickup or
induction caused by nearby
electric motor.
f) Broken
probe.
a)
Fill probe with electrolyte &
replace membrane. See pages
49 - 50.
b)
Tap probe to remove bubbles.
c)
Clean the probe and re-
calibrate.
d)
Make sure sample entirely
covers the probe sensors.
e)
Move or switch off interfering
motor.
f) Replace
probe.
“OR” on upper display
a) Probe is shorted.
b) Probe is in an out-of-range
solution.
c) Broken
probe.
a)
Test probe. Make sure probe is
fully connected to meter.
b)
Use different solution.
c) Replace
probe.
See
“Accessories” on Page 55
Temperature reading
erratic or lower
display reads “OR”
a)
Temperature of solution is out
of range.
a)
Heat or cool solution.
Slow response
a)
Dirty / Oily probe.
a)
Clean probe. See “Probe Care
& Maintenance”, page 46 – 47.