8. Troubleshooting
Cause
Problem
Solution
Power on, but no display
a)
Loose connections
a)
Ensure cables make good
contact
b)
Incorrect output voltage of
the power adaptor
b)
Use an power adaptor with
specified output voltage
Unstable pH reading
a)
Dirty electrode
a)
Clean electrode and recalibrate
b)
Electrical noise interference
b)
Switch to Symmetric mode
a)
Electrical noise interference
Oscillating temperature
readings
a)
Ensure shield wire is properly
connected to pin 7
Slow response
a) Dirty / Oily electrode
a) Clean electrode
Blinking ATC
a)
No temperature probe
connection during ATC mode
a)
Ensure temperature sensing
cable makes good contact
a)
Error in calibration
Blinking electrode
annunciator
a)
Ensure calibration standard
solution is not contaminated.
Ensure electrode is clean.
Or (pH)
a)
pH electrode is not
connected
a)
Ensure electrode makes good
contact with Transmitter
Or (Temperature)
a)
Temperature probe is not
connected when ATC
enabled
a)
Ensure electrode makes good
contact with Transmitter