5.5 Configuration Settings
Configuration settings let you configure the pH Monitor to different measurement &
operation modes & reset the pH Monitor to factory defaults.
Asymmetrical Operation Mode (ASY):
This is the default mode of operation of the pH Monitor, which is suitable for most
applications where sample (or process) liquid is not interfered by adjacent electrical
noise.
Symmetrical Operation Mode (SY):
This is mode of operation is recommended for applications where sample (or process)
liquid is subjected to high electrical interference and therefore the pH reading
becomes unstable. When the pH Monitor is set to symmetrical mode, you should
immerse the potential matching pin in the sample (or process) liquid or use a pH
electrode with built-in potential matching pin. The ‘
MEAS
’ indicator blinks when the
pH Monitor is set to Symmetric mode.
ENT
SETUP
SETUP
ENT
SETUP
SETUP
SETUP
ENT
ENT
ENT
SETUP
ENT
ENT
SETUP
SETUP
MEAS
pH
°C
ATC
READY
1
2
2
3
3
4
4
4
From pH (or ORP) measurement mode
press ENT key
to enter setup
mode as described in section
1
1
5.1. The LCD shows the first screen of
setup mode (
OFS
).
Press
▲
or
▼
key
to
select configuration settings
screen (
Cnfg
).
Press ENT key
to access configuration settings (
Cnfg
).
Select measurement mode:
The upper display shows the last
configured measurement mode (‘
PH
’ or ‘
OrP
’).
2
Press
▲
or
▼
key
to select your desired measurement mode.