EUTECH INSTRUMENTS EC-ISA-PB1-BT Instruction Manual Download Page 9

Instruction Manual 

 

Lead Electrode

 

 

7

Direct Measurement of Lead (using an ion meter) 

 

1. 

By serial dilution of the 0.1M or l,000 ppm lead standard, prepare two lead standards 
whose concentration is near the expected sample concentration. Measure out 50 ml of each 
standard into individual 150 ml beakers. Add 50 ml of methanol-formaldehyde solution 
and 2 ml of ISA to each beaker. 

2. 

Place the more dilute solution on the magnetic stirrer and begin stirring at a constant rate. 
Assure that the meter is in the concentration mode. 

 
3. 

Lower the electrode tips into the solution. 

 
4. 

Adjust the meter to the concentration of the lead standard and fix the value in the memory 
according to the meter manufacturer's instructions after stabilization of the reading. 

 
5. 

Rinse the electrodes with distilled water and blot dry. 

 
6. 

Place the more concentrated solution on the magnetic stirrer and begin stirring at a 
constant rate. 

 
7. 

Lower the electrode tips into the solution. 

 
8. 

Adjust the meter to the concentration of the lead standard and fix the value in the memory 
according to the manufacturer's instructions after stabilization of the reading. 

  
9. 

Place 50 ml of the sample, 50 ml of methanol-formaldehyde solution, and 2 ml of ISA into 
a 150 ml beaker. Place the beaker on the magnetic stirrer and begin stirring at a constant 
rate. 

 
10. 

After rinsing the electrodes, blot dry, and lower the electrode tips into the solution.  After 
stabilization, read the concentration directly from the meter display. 

 
11.  The calibration should be checked every two hours. Assuming no change in ambient 

temperature, immerse the electrode tips in the more dilute standard. After the reading has 
stabilized, compare it to the original reading recorded in Step 4 above. A reading differing 
by more than 0.5 units or a change in ambient temperature will necessitate repetition of 
Steps 2-8 (2-9) above. The meter should be re-calibrated daily. 

 

Summary of Contents for EC-ISA-PB1-BT

Page 1: ...EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL LEAD ION ELECTRODE...

Page 2: ...Lead using an ion meter 7 Low Level Lead Determination using a pH mV meter 8 Low Level Lead Determination using an ion meter 9 Titration 10 Titration of Lead 10 Titration of Sulfate 11 ELECTRODE CHAR...

Page 3: ...nations using a pH mV meter ERROR BOOKMARK NOT DEFINED TITRATION Error Bookmark not defined Titration of Sulfide ERROR BOOKMARK NOT DEFINED Titration of Silver ERROR BOOKMARK NOT DEFINED Low Level Chl...

Page 4: ...solution and standard preparation 2 Methanol formaldehyde solution To prepare this solution from your own laboratory stock add 3 drops of 37 formaldehyde to 1 000 ml of reagent grade methanol This sol...

Page 5: ...l of methanol formaldehyde solution and 50 ml of distilled water Add 2 ml of ISA Place the beaker on a magnetic stirrer and begin stirring at a constant rate After assuring that the meter is in the mi...

Page 6: ...n temperature To counteract this effect place a piece of insulation material such as styrofoam sheet between the stirrer and beaker Use plastic lab ware for all low level measurements in order to mini...

Page 7: ...ter 1 By serial dilution prepare 10 2 M 10 3 M and 10 4 M or 100 ppm and 10 ppm standards from the 0 1M or 1 000 ppm standards Prepare standards with a composition similar to the samples if the sample...

Page 8: ...ar portion of the curve requires the use of low level procedures 6 To a clean dry 150 ml beaker add 50 ml of the sample 50 ml of methanol formaldehyde solution and 2 ml of ISA Place the beaker on the...

Page 9: ...on the magnetic stirrer and begin stirring at a constant rate 7 Lower the electrode tips into the solution 8 Adjust the meter to the concentration of the lead standard and fix the value in the memory...

Page 10: ...a 150 ml plastic beaker Place the beaker on the magnetic stirrer and begin stirring at a constant rate 4 Place the electrode tips in the solution Assure that the meter is in the mV mode 5 Add incremen...

Page 11: ...low level ISA to a 150 ml plastic beaker Place the beaker on the magnetic stirrer and begin stirring at a constant rate Lower the electrode tips into the solution When the reading has stabilized fix t...

Page 12: ...lask gently to dissolve the solid Fill the flask to the mark with distilled water cap and upend the flask several times to mix the solution 2 Fill a 50 ml buret with the EDTA solution Pipet 50 ml of t...

Page 13: ...re ideal for this titration method If the sulfate samples are more dilute the lead perchlorate titrant should be correspondingly more dilute 1 Prepare 0 01M lead perchlorate titrant by pipeting 100 ml...

Page 14: ...m the unknown is calculated as follows Vt Mt MSO4 2 VSO4 2 where MSO4 2 concentration of sulfate ion in the unknown moles liter Vt volume of lead added at endpoint Mt lead concentration moles liter VS...

Page 15: ...y pH adjustment to above pH 4 by the addition of NaOH Precipitation and Complexation Sulfide phosphate hydroxide and other ions precipitate insoluble lead salts The level of lead ion the level of the...

Page 16: ...dy with one hand bring the membrane of the electrode down perpendicular to the paper and with a slight swirling motion gently polish the tip of the electrode against the surface of the polishing paper...

Page 17: ...Electrode Life A lead electrode will last six months in normal laboratory use On line measurements might shorten operational lifetime to several months In time the response time will increase and the...

Page 18: ...t interfere with the electrode s response to lead ions The reference electrode must also be considered When two solutions of different composition are brought into contact with one another liquid junc...

Page 19: ...cted substitute another lead ion electrode that is known to be in good working order for the questionable electrode If the problem persists and you are using an electrode pair try the same routine wit...

Page 20: ...e analysis method is clearly understood and is compatible with the sample Refer to the instruction manual again Reread sections on GENERAL PREPARATION and ELECTRODE CHARACTERISTICS If trouble still pe...

Page 21: ...with shorting strap Unstable Readings readings air bubble on membrane remove bubble by re dipping electrode continuously or rapidly changing electrode exposed soak electrode in lead standard to interf...

Page 22: ...correct scaling plot millivolts on the linear axis On but calibration of semi log paper the log axis be sure concentration curve is good numbers within each decade are increasing with increasing conce...

Page 23: ...us solutions only no organic solvents Size 110 mm length 12 mm diameter 1 m cable length Storage Store in lead solution ORDERING INFORMATION CODE NO DESCRIPTION EC PB 03 Lead Ion Electrode epoxy body...

Reviews: