EUTECH INSTRUMENTS EC-ISA-KO1-BT Instruction Manual Download Page 6

Instruction Manual 

 

Potassium Electrode 

 

6

All samples and standards must be aqueous. They must not contain organic solvents. Interferences 
found in Table 3 should be absent. 
 

Units of Measurement 

 
Potassium concentrations are measured in units of parts per million as potassium, parts per million 
as KCl, moles per liter, or any other convenient concentration unit. Table 1 indicates some of the 
concentration units. 
 
    TABLE 1:  Concentration Unit Conversion Factors 
 
         ppm K

+1

    

ppm KCl       moles/liter K

+1 

           3.91 

  

 7.46   

1x10

-4

 

          39.10 

  

 74.60   

1x10

-3

 

         391.00   

746.00  

1x10

-2 

 
 

Measurement Procedure 

 

Direct Measurement 

 
Direct measurement is a simple procedure for measuring a large number of samples. A single meter 
reading is all that is required for each sample. The ionic strength of samples and standards should 
be made the same by adjustment with ISA for all potassium samples. The temperature of both 
sample solution and of standard solutions should be the same. 
 

Direct Measurement of Potassium (using a standard pH/mV meter) 

 

1. Prepare 

10

-2

, 10

-3

, and 10

-4

M or 100, 10, and 1 ppm standards by serial dilution of the 

0.1M or 1,000 ppm standard. Measure out 100 ml of each standard into individual 150 ml 
beakers.  Add 2 ml of ISA per 100 ml of standard. 

 
2. 

Place the most dilute solution (10

-4

M or 1 ppm) on the magnetic stirrer and begin stirring 

at a  constant rate. After assuring that the meter is in the mV mode, lower the electrode 
tip(s) into the solution. When the reading has stabilized, record the mV reading. 

 
3. 

Place the midrange solution (10

-3

M or 10 ppm) on the magnetic stirrer and begin stirring.  

After rinsing the electrode(s) with distilled water and blotting dry, immerse the electrode 
tip(s) in the solution.  When the reading has stabilized, record the mV reading. 

 
4. 

Place the most concentrated solution (10

-2

M or 100 ppm) on the magnetic stirrer and begin 

stirring. After rinsing the electrode(s) with distilled water and blotting dry, immerse the 
electrode tip(s) in the solution.  When the reading has stabilized, record the mV reading. 

5. 

Using the semi-logarithmic graph paper, plot the mV reading (linear axis) against the 
concentration (log axis). Extrapolate the curve down to about 1x10

-5

M or 0.2 ppm. A 

typical calibration curve can be found in Figure 1. 

 
 
 
 

Summary of Contents for EC-ISA-KO1-BT

Page 1: ...Instruction Manual Potassium Electrode 1 EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL POTASSIUM ION ELECTRODE...

Page 2: ...of Potassium using a standard pH mV meter 6 Direct Measurement of Potassium using an ion meter 7 Low Level Potassium Determination using a standard pH mV meter 8 Low Level Potassium Determination usi...

Page 3: ...grade potassium chloride Swirl the flask gently to dissolve the solid Fill to the mark with distilled water cap and upend several times to mix the solution 3 Eutech Potassium Standard 1 000 ppm K 1 Co...

Page 4: ...instability is observed see the TROUBLESHOOTING section 2 Using a pipet add 1 ml of 0 1M or 1 000 ppm potassium standard to the beaker When the reading is stable record the mV reading 3 Using a pipet...

Page 5: ...perature To counteract this effect place a piece of insulating material such as styrofoam sheet or asbestos sheet between the stirrer and beaker Always rinse the electrode tip s with distilled water a...

Page 6: ...2 10 3 and 10 4 M or 100 10 and 1 ppm standards by serial dilution of the 0 1M or 1 000 ppm standard Measure out 100 ml of each standard into individual 150 ml beakers Add 2 ml of ISA per 100 ml of st...

Page 7: ...de tip s in the solution When the reading has stabilized record the millivolt reading Determine the concentration directly from the calibration curve 7 The electrode s should be re calibrated every 1...

Page 8: ...meter display 12 The electrodes should be re calibrated every 1 2 hours Simply repeat Steps 2 8 2 9 above Low Level Potassium Determination using a standard pH mV meter This procedure is recommended f...

Page 9: ...tip s into the solution 10 After the reading has stabilized record the mV reading and determine the concentration from the low level calibration curve 11 Prepare a new low level calibration curve dai...

Page 10: ...ris hydroxymethyl aminomethane Interferences ppm 100 ppm K 1 10 ppm K 1 1 ppm K 1 Cs 1 1 0X102 1 0X101 1 0 K 1 2 7X102 2 7X101 2 7 Tl 1 3 1X103 3 1X102 3 1X101 H 1 1 6 pH 2 6 pH 3 6 pH Ag 1 2 7X105 2...

Page 11: ...chloride solutions The upper limit of detection is above 1 0X10 1 M when other ions are present but the possibility of a liquid junction potential developing at the reference electrode and the salt e...

Page 12: ...rained of filling solution if refillable and the rubber insert placed over the filling hole ELECTRODE THEORY Electrode Operation A potassium ion electrode consists of an electrode body containing an i...

Page 13: ...difference between the two solutions since ions diffuse at different rates When making measurements it is important to remember that this potential be the same when the reference is in the standardiz...

Page 14: ...es in the sample may be present or the technique may be faulty See Reagents Sample and Technique sections below 5 If another electrode is not available for test purposes or if the electrode in use is...

Page 15: ...Reading see meter instruction manual electrode s not unplug electrode s and reset plugged in properly reference electrode be sure reference electrode is not filled filled air bubbles on remove bubble...

Page 16: ...se recommended ISA standard used as ISA use ISA defective electrode check electrode operation electrode exposed soak electrode standard to interferences in potassium air bubble on membrane remove bubb...

Page 17: ...ducibility 2 Samples Aqueous solutions only no organic solvents Size 110 mm length 12 mm diameter 1 m cable length Storage Electrode should be stored in dilute potassium standard ORDERING INFORMATION...

Reviews: