EUTECH INSTRUMENTS EC-ISA-KO1-BT Instruction Manual Download Page 12

Instruction Manual 

 

Potassium Electrode 

 

12

 

pH Effects 

 
Hydrogen ion interferes with measurements of low levels of potassium ion although the electrode 
can be used over a wide pH range. Table 3 should be used to determine the minimum pH at which 
low level potassium measurements can be made without more than a 10% error due to hydrogen ion 
interference. 

 

Electrode Life 

 
A potassium electrode will last six months in normal laboratory use. On-line measurement might 
shorten operational lifetime to several months. In time, the response time will increase and the 
calibration slope will decrease to the point calibration is difficult and electrode replacement is 
required. 
 

Electrode Storage 

 
The electrode may be stored in 1.0X10

-2

M potassium standard for short periods of time. For storage 

over 3 weeks, rinse and dry the potassium membrane electrode and cover the tip with any protective 
cap shipped with the electrode(s). The reference portion of the combination electrode (or the outer 
chamber of the reference electrode) should be drained of filling solution, if refillable, and the rubber 
insert placed over the filling hole. 
 

ELECTRODE THEORY 

 

Electrode Operation 

 
A potassium ion electrode consists of an electrode body containing an ion exchanger in a sensing 
module. This sensing module contains a liquid internal filling solution in contact with a gelled 
organophilic membrane containing a potassium selective ion exchanger. 
 
An electrode potential develops across the membrane when the membrane is in contact with an 
potassium solution. Measurement of this potential against a constant reference potential with a 
digital pH/mV meter or with a specific ion meter depends on the level of free potassium ion in 
solution. The level of potassium ions, corresponding to the measured potential, is described by the 
Nernst equation: 
 
                    E = Eo + S log X 
       where: 
                    E = measured electrode potential  
                    Eo= reference potential (a constant) 
                    S = electrode slope (~ 56 mV/decade) 
                    X = level of potassium ions in solution 
 
The activity, X, represents the effective concentration of the ions in solution. Total potassium 
concentration, Ct, includes free potassium ions, Cf, plus bound or complexed potassium ions, Cb. 
Since the potassium electrodes only respond to free ion, the free ion concentration is: 
 
 

 

 

 

Cf = Ct - Cb 

 

Summary of Contents for EC-ISA-KO1-BT

Page 1: ...Instruction Manual Potassium Electrode 1 EUTECH INSTRUMENTS PTE LTD INSTRUCTION MANUAL POTASSIUM ION ELECTRODE...

Page 2: ...of Potassium using a standard pH mV meter 6 Direct Measurement of Potassium using an ion meter 7 Low Level Potassium Determination using a standard pH mV meter 8 Low Level Potassium Determination usi...

Page 3: ...grade potassium chloride Swirl the flask gently to dissolve the solid Fill to the mark with distilled water cap and upend several times to mix the solution 3 Eutech Potassium Standard 1 000 ppm K 1 Co...

Page 4: ...instability is observed see the TROUBLESHOOTING section 2 Using a pipet add 1 ml of 0 1M or 1 000 ppm potassium standard to the beaker When the reading is stable record the mV reading 3 Using a pipet...

Page 5: ...perature To counteract this effect place a piece of insulating material such as styrofoam sheet or asbestos sheet between the stirrer and beaker Always rinse the electrode tip s with distilled water a...

Page 6: ...2 10 3 and 10 4 M or 100 10 and 1 ppm standards by serial dilution of the 0 1M or 1 000 ppm standard Measure out 100 ml of each standard into individual 150 ml beakers Add 2 ml of ISA per 100 ml of st...

Page 7: ...de tip s in the solution When the reading has stabilized record the millivolt reading Determine the concentration directly from the calibration curve 7 The electrode s should be re calibrated every 1...

Page 8: ...meter display 12 The electrodes should be re calibrated every 1 2 hours Simply repeat Steps 2 8 2 9 above Low Level Potassium Determination using a standard pH mV meter This procedure is recommended f...

Page 9: ...tip s into the solution 10 After the reading has stabilized record the mV reading and determine the concentration from the low level calibration curve 11 Prepare a new low level calibration curve dai...

Page 10: ...ris hydroxymethyl aminomethane Interferences ppm 100 ppm K 1 10 ppm K 1 1 ppm K 1 Cs 1 1 0X102 1 0X101 1 0 K 1 2 7X102 2 7X101 2 7 Tl 1 3 1X103 3 1X102 3 1X101 H 1 1 6 pH 2 6 pH 3 6 pH Ag 1 2 7X105 2...

Page 11: ...chloride solutions The upper limit of detection is above 1 0X10 1 M when other ions are present but the possibility of a liquid junction potential developing at the reference electrode and the salt e...

Page 12: ...rained of filling solution if refillable and the rubber insert placed over the filling hole ELECTRODE THEORY Electrode Operation A potassium ion electrode consists of an electrode body containing an i...

Page 13: ...difference between the two solutions since ions diffuse at different rates When making measurements it is important to remember that this potential be the same when the reference is in the standardiz...

Page 14: ...es in the sample may be present or the technique may be faulty See Reagents Sample and Technique sections below 5 If another electrode is not available for test purposes or if the electrode in use is...

Page 15: ...Reading see meter instruction manual electrode s not unplug electrode s and reset plugged in properly reference electrode be sure reference electrode is not filled filled air bubbles on remove bubble...

Page 16: ...se recommended ISA standard used as ISA use ISA defective electrode check electrode operation electrode exposed soak electrode standard to interferences in potassium air bubble on membrane remove bubb...

Page 17: ...ducibility 2 Samples Aqueous solutions only no organic solvents Size 110 mm length 12 mm diameter 1 m cable length Storage Electrode should be stored in dilute potassium standard ORDERING INFORMATION...

Reviews: