Instruction Manual
CyberScan pH1500
13
4.1.1 P1.0 View pH Electrode Offset
This setup menu allows you to view the pH electrode offset in mV
after 1-point standardization is successfully done.
pH
View
press
to set value
press
to
accept
enter
ELECTRODE
To view pH electrode offset
1. Access View the pH electrode offset menu from pH measurement
screen by pressing setup key
2. Press
▲
and
▼
keys to scroll until display shown as above.
save
view
3. Press enter to view the offset. A series of dashes “- - -“ will be
displayed if the unit has not been standardized.
pH
mV