3.5 Manual
Calibration
In Manual Calibration mode you are not limited to the conductivity calibration standards
listed previously in Table 1. This example shows a manual calibration sequence using a
12.00 mS/cm conductivity calibration standard.
1.
If necessary, press
MODE
key to select conductivity
mode.
1.
Rinse the probe thoroughly with de-ionized water or
a rinse solution, then rinse with a small amount of
calibration standard.
1.
Dip the probe into the calibration standard. Stir the
probe gently to create a homogeneous sample.
Allow time for the reading to stabilize.
1. Press
CAL
to enter conductivity calibration mode.
The [
CA
] indicator will appear briefly, then a value
will appear flashing.
NOTE: To exit calibration without confirmation, press
CAL to return to measurement mode.
2.
When the value is stable, press
▲
or
▼
as needed
to adjust the value to match your calibration
standard.
3. Press
ENTER
to confirm the adjusted value. [
CO
]
will appear briefly indicating that the calibration was
successful. The meter returns to measurement
mode.
4.
Repeat steps 1-6 as needed with additional
calibration standards.
mS
mS
mS
mS
CAL
HOLD
ENTER
Measurement Mode