Instruction Manual
COND 6+, TDS 6+, SALT 6+
-11-
With manual calibration, you are not limited to the calibration standard values listed
previously. Manual calibration is useful when you wish to use one or more standard
values that are not listed above.
See Section
5.3 Advanced Setup
to modify automatic or manual calibration.
3.4
Using Automatic Calibration (COND 6+)
In Automatic Calibration mode, the COND 6+ can accept up to 4 calibration points with
maximum of 1 point per measurement range. Note: values in the 0.00 to 20.00 µS/cm
range cannot be calibrated in Auto Calibration mode.
1.
If necessary, press
MODE
key to select conductivity
mode.
2.
Rinse the probe with deionized water or a rinse
solution, then rinse with a small amount of calibration
standard.
NOTE: For Automatic Calibration you must use one of
the calibration standards listed in Table 1.
3.
Dip the probe into the calibration standard. Stir the
probe gently to create a homogeneous sample. Allow
time for the reading to stabilize.
4. Press
CAL
key to enter conductivity calibration mode.
The [
CA
] indicator will appear briefly, then a value will
appear flashing.
NOTE: To exit calibration without confirmation, press
CAL to return to measurement mode.
5.
When the value is stable, press
ENTER
. The
calibration standard value will appear for 3 seconds.
If the calibration is successfully performed, [
donE
] will
be displayed briefly before meter returns to
measurement mode.
6.
Repeat steps 1-5 as needed with additional calibration
standards.
mS
mS
mS
mS
CAL
HOLD
ENTER
Measurement Mode