26
10
SPECIFICATIONS
Conductivity Range
Resolution
Cell Constant
0.00 – 20.00 µS/cm
0.01 µS/cm
0.1
0.0 – 200.0 µS/cm
0.1 µS/cm
0.1
0.0 – 200.0 µS/cm
0.1 µS/cm
1
0 – 2000 µS/cm
1 µS/cm
1
0.00 – 10.00 mS/cm
0.01 mS/cm
1
0.00 – 20.00 mS/cm
0.01 mS/cm
1
0.0 – 200.0 mS/cm
0.1 mS/cm
1
0.0 – 200.0 mS/cm
0.1 mS/cm
10
Relative Accuracy
± 1 % of full scale
Temperature
-10 to 110
o
C
Resolution/Accuracy
0.1 / ± 0.5
o
C
Sensor
Pt 100 (3-wire)
Temperature Compensation
Automatic / Manual (0 to 50
°
C)
Temperature Coefficient
0.0 to 5.0 %
Set-point And Controller Functions
Function Limit
Control
Switching Conductivity Hysteresis
5% of full scale
Contact Outputs, Controller
2 SPDT relays
Switching Voltage / Current / Power
Max 250 VAC / Max 3A / Max 600 VA
Signal Output / Load
4-20 mA galvanically isolated / 200
Ω
Electrical Data And Connections
Power Requirements
85 to 260 VAC, 50/60 Hz or DC
Rated Input Current (MAX)
100 mA
Insulation Category
II
Connection Terminals
3 Detachable connectors
(3-pin; 7-pin & 9-pin terminal blocks) +
1 FIXED CONNECTOR
Main Fuse
250 mA, Anti-surge (BUSSMAN S504+250mA)
Environmental Conditions
Ambient Temp. Operating Range
- 10 to 50
o
C (14 to 122
o
F)
Rel. Humidity
10 to 95 % (non-condensing)
Mechanical Specifications
Dimensions (Panel Housing – W x H x D)
96 x 48 x 150 mm
Weight
300g (350g boxed)