background image

 

11 

5

 

CONDUCTIVITY CALIBRATION 

IMPORTANT

: When Calibration mode is entered, controller automatically goes into 

a “HOLD” mode where the 4-20 mA output freezes and relays are de-activated (if it 
was in an activated condition). Upon return to measurement mode, both 4-20mA 
output and relay activities resume, depending on settings. 
 
The Alpha COND 200 controller includes 8 conductivity measurement ranges. 
One-point slope calibration is possible, in each range.  
Choose fresh standard solutions, whose value is reasonably close to the 
measurement value. 
Before calibration, make sure electrode is clean. Use iso-propyl alcohol followed by 
thorough rinsing in distilled water to clean electrode. 
 

Important

: To achieve a successful conductivity calibration, two conditions must 

be satisfied: 

1.

 

Difference between measured value of standard solution, and actual 
value of standard solution, must be within ± 40 %. 
(e.g. if value of standard solution is 1413 uS/cm, then allowable 
measured values by controller must be within 1978.2 uS/cm {+ 40%} 
and 847.8 uS/cm {- 40%}); and 

2.

 

the lowest possible value must be at least 10% of full scale 
(e.g. if range of 0 to 200.0 uS/cm selected, min. measured value at 
calibration, must be at least 20.0 uS/cm). 

 

If any or both conditions are not satisfied, controller will display ERR 1 (blinking). 
 
1.

 

Key in the password “011” using the method described in Section 5. 

2.

 

Controller displays the value of the standard solution.  

3.

 

Use electrode to agitate standard solution to obtain a homogeneous solution 
and to dislodge any bubbles. Allow electrode time to stabilize with solution 
temperature. 

4.

 

Use 

keys to adjust displayed readings to the value of the standard 

solution. Press ENTER. 

5.

 

If any of the two conditions mentioned above is not satisfied, controller will 
display ERR1, blinking. 

6.

 

If calibration is successful, Controller displays DONE, blinking. Press 
ENTER to revert to Conductivity measurement mode (Relays and 4-20 mA 
output resume previous settings). 

NOTE: To clear the ERR1 display and return to calibration mode, press 

 

keys together. 

Summary of Contents for Alpha COND 200

Page 1: ...Instruction Manual Alpha COND 200 1 8 DIN Conductivity Controller Transmitter S mS C 68X276103 Rev 3 Apr 2008...

Page 2: ......

Page 3: ...troller transmitter as possible If there are doubts in the use of the Alpha COND 200 controller transmitter do not hesitate to contact your nearest lpha Authorized Distributor The information presente...

Page 4: ...RATURE CALIBRATION 13 7 SETUP MODE 14 7 1 GENERAL INFORMATION 14 7 2 SETUP MODE OVERVIEW 15 7 3 SET POINT 1 P1 0 16 7 4 SET POINT 2 P2 0 18 7 5 MEASUREMENT RANGE SELECTION P3 0 20 7 6 CONFIGURE TEMPER...

Page 5: ...set point two SPDT relay operation Scrolling 14 segment LED guides user easily through setup functions Reliable power supply from 85 to 260 V AC 50 60 Hz or DC withstands voltage fluctuations Push but...

Page 6: ...that power source it is connected to is as specified in the top label Failure to do so may result in a permanent damage to the unit The unit has live and exposed parts inside If it has to be opened ma...

Page 7: ...e in micro Siemens cm 7 oC Unit of the displayed parameter temperature 8 mS Displayed when measurement range in milli Siemens cm Keys 9 MODE Toggle between Conductivity and temperature display 10 ENTE...

Page 8: ...position normally closed 6 Relay 2 center pole 7 Relay 2 activated position normally open 8 Relay 1 deactivated position normally closed 9 Relay 1 center pole 10 Relay 1 activated position normally op...

Page 9: ...ads to terminals 13 to 15 of the seven pin terminal block Either wire can be connected to either terminal Terminals 13 and 14 must be shunted unless using a 3 wire RTD NOTE CON 200 is factory set for...

Page 10: ...to the controller 1 Align the catch to the side of the controller and insert threaded rods through catch 2 Screw the threaded rod through the catch in a clock wise direction Tighten until the catch h...

Page 11: ...stalling Split Ferrite Clamping The power cable L N E need to be connected to the instrument with two turns through Split Ferrite Wurth Electronika clamp which is supplied as an accessory with the ins...

Page 12: ...mS cm 5 0 01 mS cm 1 0 00 20 00 mS cm 6 0 01 mS cm 1 0 0 200 0 mS cm 7 0 1 mS cm 1 0 0 200 0 mS cm 8 0 1 mS cm 10 Temperature Measurement mode Press MODE key once to view the temperature measurement...

Page 13: ...reads P 000 The first 0 flashes 2 Press ENTER again to leave the first digit 0 and to scroll to the next number 3 Press the or keys to change the second digit to the correct number 1 or 2 Press ENTER...

Page 14: ...10 P P 000 P 000 P 0 10 P 020 P 011 P 022...

Page 15: ...olution is 1413 uS cm then allowable measured values by controller must be within 1978 2 uS cm 40 and 847 8 uS cm 40 and 2 the lowest possible value must be at least 10 of full scale e g if range of 0...

Page 16: ...up program See Setup program P5 0 Controller displays calibrated conductivity point for selected range If calibration is not done for the selected range controller displays If after thorough cleaning...

Page 17: ...reen will flash in conductivity mode and an error message OR is displayed in temperature mode MTC MTC lets you select a specific value at which temperature will be compensated You can select a manual...

Page 18: ...ement mode both 4 20mA output and relay activities resume depending on settings To enter setup mode 1 Key in password 022 using method described in section 5 2 Press or keys to display various sub men...

Page 19: ...with the corresponding cell constant value Eight measurement ranges are available P4 0 Temperature Data P4 1 select ATC or MTC P4 2 select temperature coefficient value P4 3 select normalization tempe...

Page 20: ...value 4 Press or keys and adjust first relay set point Default value is 10 of full scale of range selected 5 Press ENTER to confirm and continue to step three of P1 2 or press and keys together to ret...

Page 21: ...ot de activate till measured Conductivity value rises above 300 0 S cm Default hysteresis value is 5 of full scale The hysteresis window can be set to any value within the range shown below 1 Key in p...

Page 22: ...point value 4 Press or keys and adjust second relay set point Default value is 90 of full scale of range selected 5 Press ENTER to confirm and continue to step three of P2 2 or press and keys togethe...

Page 23: ...ll not de activate till measured Conductivity value drops below 1700 0 S cm Default hysteresis value is 5 of full scale The hysteresis window can be set to any value within the range shown below 1 Key...

Page 24: ...GE Press ENTER 3 The screen will scroll P3 1 and the selected range 4 Press or keys to select appropriate range and cell Available ranges are as follows Default range is highlighted Conductivity Range...

Page 25: ...and keys together again to return to Conductivity measurement mode P4 2 Setting Temperature Coefficient 1 Key in the password 022 as per procedure in Section 5 2 Press key until screen displays P4 0 a...

Page 26: ...change each time you recalibrate the controller P 5 0 C L P 5 1 CA L 300 Press and keys together ESCAPE at anytime to leave the Setup mode and return to Measurement mode P5 0 Viewing Calibrated Condu...

Page 27: ...P6 1 Viewing Conductivity electrode data 1 Key in the password 022 as per procedure in Section 5 2 Press key until screen displays P6 0 and ELEC Press ENTER 3 Screen will scroll P6 1 FCTR then cell c...

Page 28: ...et to 25 oC Temperature Coefficient is reset to 2 1 Normalization Temperature is reset to 25 0 oC NOTE ATC temperature offset is NOT reset P 7 0 RsT n 0 y e S P 7 1 Press and keys together ESCAPE at a...

Page 29: ...urrent will be proportional to the Conductivity value displayed on the panel and according to the measurement range selected Conductivity Range Range No R Cell Constant 4 mA current 20 mA current 0 00...

Page 30: ...nction Limit Control Switching Conductivity Hysteresis 5 of full scale Contact Outputs Controller 2 SPDT relays Switching Voltage Current Power Max 250 VAC Max 3A Max 600 VA Signal Output Load 4 20 mA...

Page 31: ...e sensor w 3 wire Pt100 k 1 0 ECCONSEN88X Note Above Conductivity electrodes withstand up to 3 bar pressure These cells have integral 1 0 m 5 wire double shielded open ended cable Please contact your...

Page 32: ...owing settings will remain unchanged 1 Measurement range 2 Temperature compensation mode ATC or MTC 3 Temperature offset calibration value if in ATC mode Conductivity Defaults Conductivity range Remai...

Page 33: ...control element provided its power requirements does not exceed this However to preserve the life of your controller or if higher power is needed it is recommended that you use the controller relay t...

Page 34: ...of Goods Authorisation number before returning items for any reason When applying for authorisation please include data requiring the reason of return Items must be carefully packed to prevent damage...

Page 35: ......

Page 36: ...nstruments Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949 Tel 65 6778 6876 Fax 65 6773 0836 E mail eutech thermofisher com Web site http www thermo com process Oakton Instruments P O Box 5136 Ve...

Reviews: