5.5 Setting Temperature Coefficient
5.5.1 Temperature Coefficient
This menu allows you to set the Temperature Coefficient
corresponding to the solution whose Conductivity is being
measured. In most controllers, this is fixed at 2.10% per
o
C.
However, for alpha CON 100 C/CX models, it is adjustable
from 0.0 to 10.0% per
o
C (please refer to Appendix 3 for
Temperature Coefficient determination and limits).
1. Enter this menu by pressing
the
MODE
key to the
“SEt”
display if you are in the
measurement mode.
2.
Press
ENTER
key to access
the SETUP menu. Press
▲
key
twice to enter the “
SEt tc
” menu. If
you do not wish to effect any
change and keep the Temperature
Compensation as the previous
value, press
MODE
key. This will
take you out of this mode and back
to its measurement mode.
3. Press
ENTER
and the display
shows the current value of
Temperature Coefficient (blinking).
4. Use
▲
or
▼
key to change the
value of the Temperature Coefficient. Press
ENTER
to
confirm the value to be used. Press
MODE
key to exit
to measurement mode.
SETUP
MODE
ENTER
tc
SETUP
MODE
ENTER
2.10
SETUP
MODE
ENTER
SEt