10-2. Measuring operation
Aim towards the forehead (see the diagram below for the positioning), from a distance of 1-3cm (0.4-
1.2 in), press the measuring trigger, the temperature is displayed Immediately. Making sure there is no
hair, perspiration, cosmetic or cap covered on the forehead.
If it can't measure the forehead temperature due to perspiration or other factors, please take
the
temperature behind the ear lobe, make sure there is no hair, perspiration, cosmetic or cap
covered.
10-2-1. The ambient temperature should be stable, don't measure the body temperature in front of
fans, air outlets of the air-condition etc. where the airflow is floating. 10-2-2. To obtain stable and
reliable data, you should keep the device for 15-20 minutes to balance its own temperature before
measuring when the ambient temperature change greatly, and keep it 1 minute after 2 continuous
measuring.
10-2-3. Please don't take the body temperature if the patient is sweating or have done the cold
compresses etc., because they will lower the measuring result.
10-2-4. Suggest measure the body temperature three times and be subject to the highest value.
10-3. Buttons operation
10-3-1. ON/SCAN button (Measuring Trigger) Press the button to power on the device
and begin to measure.
10-3-2. MODE button: Switch the body mode, surface mode.