After measurement, you can review the results in the memory bank by pressing the “MEM”
button. Alternatively, you can press “MEM” button in Off Mode to display the stored results.
If there are results in the memory bank, the LCD will display the most recent result including
the memory number and blood pressure.
9.
Troubleshooting
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD shows low battery
symbol
Low Battery
Change the batteries
LCD shows “Er 0”
Pressure system is unstable
before measurement
Don’t move and try again.
LCD shows “Er 1”
Fail to detect systolic pressure
LCD shows “Er 2”
Fail to detect diastolic pressure
LCD shows “Er 3”
Pneumatic system blocked or
cuff is too tight during inflation
Apply the cuff correctly and try again
LCD shows “Er 4”
Pneumatic system leakage or
cuff is too loose during inflation
LCD shows “Er 5”
Cuff pressure above 300mmHg
Measure again after five minutes. If
the monitor is still abnormal, please
contact the local distributor or the
factory.
LCD shows “Er 6”
More than 3 minutes with cuff
pressure above 15 mmHg
LCD shows “Er 7”
EEPROM accessing error
LCD shows “Er 8”
Device parameter checking error
LCD shows “Er A”
Pressure sensor parameter error
No response when you press
button or load battery.
Incorrect operation or strong
electromagnetic interference.
Take out batteries for five minutes,
and then reinstall all batteries.