User manual TOM600
Item partnumber : 61-200-0600
6
Now the unit will execute a reset, and in the display appears if timer is on, else go to point 3.2 Timer off.
T=xxxs Avg Auto
(Lap) (+10V, +100V)
PRESS START !
By pressing key (A) „Start” the next measurement will be started
.
The timer decrements in 1 second steps up to Zero, then the measure value will appears in the display as
R
avg
by AVERAGE - Mode and R
lap
by Lapse – Mode.
In the 2
nd
line the air temperature and the humidity is displayed:
Ravg = 2.5xE9
Ω
(Rlab = 2.5 G
Ω
)
AIR: xx°C xx%r.F
Alternates all 2 second in the 2
nd
line appears:
FN:xxx T=xxxs
By pressing key (A) „Start“ the measure value together with the air temperature and the Humidity will be
saved in the EEPROM using the displayed File – No. These values can be read later with the PC - software.
In the meanwhile you can do a manual registering to make the correspondance between the file number and
the measured items (example : file 1 : table mat ; file 2 : ground mat…)
3.2 Timer
If the timer is not activated, the EUROSTAT 600 directly switches on the resistance display in the on-line
operation. According to the chosen display mode and voltage mode in the display appears:
Example:
R=5.5xE10
Ω
T=OFF
Ù
AUTO
AIR: xx°C xx%r.F
4.
Other Display Information’s
If the maximum measurement is exceeded appears at the display:
Measure voltage = Auto, 100V
R>2.0xE12
Ω
(R>2,0
T
Ω
)
Measure voltage = 10V
R>2.0xE11
Ω
(R>200
G
Ω
)
If the minimal measurement is under-stride appears at the display:
Measure voltage = Auto, 10V
R<2.0xE04
Ω
(R<20
k
Ω
)
Measure voltage = 100V
R<1.0xE05
Ω
(R<100
k
Ω
)
By searching the measure range, it appears in the 1
st
line of the display :
WAIT !
When the battery voltage is lower than 4,6Volt, it appears in the 2nd line of the display alternate to the actual
display-message.
LOW BATTERY !