23
Figure 3.2-5: Front panel removal
If the fuse/relay procedure does not fix the problem then replace
U1 (MC1458) and recheck the system. If the system still fails to
operate, return the unit to the factory for repair.
6. Unit reads either 0 or over 100% with sensor installed. - Sensor
is defective. Replace sensor.
If any components are replaced the system should be recalibrated
as set forth in Section 3.2.5.
3.2.7 Specifications
3.2.7.1 Model 5112
Sensor type:
Temp compensated, capacitive film
Range:
Adjustable: 0-100% RH
Accuracy:
±
3% RH
Lo/Hi band:
2% RH below set pt / 2% RH above set pt
Set pt adjust:
22-turn screw driver adjust potentiometer
Manual control:
Dehumidify ON/OFF pushbutton switch
Display:
3½-digit LCD
Resolution:
0.1% RH (
±
1 digit)
Control Out:
3 Amp solid state relay
Power req.:
90-260 VAC, 50/60Hz, 700W max
Dimensions:
3.78”Wx3.78”Hx3.35”D (DIN 43700, 9.6x9.6x8.5 cm)
3.2.7.2 Pump
Capacity (free air):
1200 cubic inch/minute
RPM:
1550
Power:
115 VAC/60 Hz, 1.5 Amps
230 VAC/50 Hz, 0.75 Amps