Model 4431T Shielded Bag Test System - Operating Manual
18
D00003 Rev. K.
Figure 7-7 – Path selection for the Oscilloscope.
Pull down the Scope menu and click on the “USB0::0x….” entry.
Pull down the Scope Menu again and click on “Refresh”
The USB devices should now be properly set up in the system. Click on the tab labeled
OPERATOR, and you will return to the Operator screen.
Figure 7-8 – Selection Switch for Monitored or Continuous
7.1.3 Select “Monitored” or “Continuous” testing. Locate the TEST MODE switch at the top of the
screen. To change the mode, click once on the switch icon. See the red arrow above.
Monitored mode
: Select MONITORED on the screen and switch the 4431 Discharge Unit selector
switch to MAN nJ. The user will be asked to insert samples and initiate each discharge by pressing the
red TEST button on the front panel of the Discharge Unit. After each waveform is captured, the user will
be asked to approve the waveform.
Continuous mode:
Select CONTINUOUS on the screen and switch the 4431 Discharge Unit selector
switch to the AUTO setting. The user will be asked to insert each sample, then the computer will
sequence through the number of discharges entered for the test. (NOTE: If the Discharge Unit is
switched to a manual setting, the computer is not controlling the discharges and will automatically run in
Monitored mode.)