12 Doc. Ref.: ETION-MED-MDF-01-06-08 (Rev 02)
Components
Interval
Activity
Breathing device
CUBE 25 NIV
When patient changed
Disinfect the device surfaces
Power cable
When patient changed
Disinfect the power cable
Coarse filter
At the latest after 1 500 operating hours or
when changing patients
Replace coarse filter,
earlier if damaged
Fine filter
Every 1 000 operating hours or when dirty
Replace fine filter
Transport bag
When patient changed
Replace transport bag
Blower
Every 20 000 operating hours
Have the device serviced at
supplier for safety check and
blower replacement.
1) Appropriate PPE must be worn when the device is decontaminated (K95 mask, face shield, gloves and
coveralls.
2) Perform cleaning in a well-ventilated area.
3) Remove coarse and fine filter.
4) Use a 3% Ready to Use RBT - Protectus Viridis solution. (Follow manufacturer instructions.)
5) Spray the outside of the device with RBT - Protectus Viridis solution.
6) Use a clean lint free cloth and wipe all obvious dirt from the device.
7) Spray the device again and wait for 5 minutes (disinfectant contact time).
8) Connect the Cube 25 NIV to ResComf XD100 (or equivalent) for a period of 30 minutes. (Follow manufacturer
instructions.)
9) Wipe off any excess disinfectant with a clean lint free cloth.
10) Replace coarse and fine filter before using it on the next patient.
CUBE 25 NIV Service, Maintenance and Repair
If the device is used without Bacterial/Viral filter or contamination is suspected,
contamination disinfection protocol shall be followed. Contact supplier for instructions.
WARNING Never use if damaged.
CUBE 25 NIV Cleaning Instructions