CUBE 25 NIV
Page 1: ...COVID 19 setup guide CUBE 25 NIV ...
Page 2: ...lifetime of fiveyears Authorization is only provided for emergency use during the COVID 19 pandemic Etion Create intends to register a CE Mark of the CUBE 25 NIV and shall inform customers as and when this will beavailable Please contact your supplier after 31 March 2021 for expanded application clinical indication information instructions for disposal or product update options Regulatory Status a...
Page 3: ... a Clinical and Patient mode Once the device is set up by a clinician the day to day operation can be performed by the Patient or Care Worker 4 Clinical and User Manual can be found here https www parsec co za downloads cube25 Feature CPAP PSV S PSV ST PCV APCV Range CPAP 3 to 20 hPa IPAP 5 to 25 hPa EPAP 3 to 23 hPa Target Volume 0 to 2000ml Backup rate 4 to 30 bpm Inspiration Trigger 1 to 5 step...
Page 4: ...I 4 Doc Ref ETION MED MDF 01 06 08 Rev02 Device Interfaces For further details refer to the User Manual Device Controls Start Stop ...
Page 5: ... used to power your Cube 25 but for backup power purposes in the event of an unexpected loss of power The UPS must be placed at least 0 5 m away from the Cube 25 and patient Be certain to plug the UPS power cord directly into a wall outlet and not into a surge protector The UPS outlets are powered whenever the UPS is switched ON During a power outage the UPS outlets will be powered for a limited t...
Page 6: ...ructed flow remove plastic rectangular filter cover 18 Connect CUBE 25 to mask 19 Connect second oxygen feed AFTER starting CUBE 25 20 Adjust flow if required on both oxygen feeds to required levels 21 Monitor patient O2 saturation FiO2 andCO2 22 Respond to alarms as necessary Audio pause will silence alarm for two minutes 23 Turn off second oxygen supply at end of therapy 24 Stop therapy on CUBE ...
Page 7: ... 9 of this document The Cube 25 NIV must be setup in the Therapy Settings of the device for Tube Length of 1 8m 22mm and the setting for the Bacterial Filter must be turned to YES The unit of measure is selectable in the Therapy Settings Selecting Therapy Modes and Therapy Sets CAUTION A physician must set the correcttherapy CUBE 25 NIV Settings for COVID 19Treatment A physician is required to sel...
Page 8: ...ap 22F with cap removed to ensure leak 1963000 CMS 4 Bacterial Viral filter in exhalation 1690001 CMS 5 Oxygen adaptor for secondary oxygen feed CCSC1963 CMS 6 Anaesthetic mask Size 3 4 or 5 1514 or 1515 or 1516 CMS 7 Standard Oxygen tubing for secondary oxygen feed 1174000 CMS 8 Uninterruptable Power Supply Not shown on image UPS850LI CMS CMS Channels Medical Solutions Mask Patient Circuit Setup ...
Page 9: ...ocument Feed 2 Secondary Oxygen feed Refer to item 7 in Mask Patient Circuit Setup on page 8 of this document NOTES This could be a bridge therapy for patients that require more than just oxygenation supplementation The oxygenation of patients should be continuously monitored Oxygen supplementation to the mask only without oxygen supply via nasal prongs seems to be not suitable for the purpose of ...
Page 10: ...dby the lettersforming each tone pitch c a f e The C is an octave above middle C Display Priority Acoustic Cause Trigger Resolution Circuit disconnection High c a f a f Fall in pressure during exhalation Tube system probably detached 10 sec Check tube connections Mains failure Moderate 2 min Beep The device no longer has power supply Direct Check the power and UPS connections Tube blocked Moderate...
Page 11: ...espiratory rate is below the set limit value 5 sec The patient s condition should be checked Target volume not reached Moderate c a f The target volume is not reached despite the maximum indicated inspiration pressure Immediately The patient s condition should be checked Target minute volume not reached Moderate c a f The target minute volume is not reached despite the maximum indicated inspiratio...
Page 12: ...eaning in a well ventilated area 3 Remove coarse and fine filter 4 Use a 3 Ready to Use RBT Protectus Viridis solution Follow manufacturer instructions 5 Spray the outside of the device with RBT Protectus Viridis solution 6 Use a clean lint free cloth and wipe all obvious dirt from the device 7 Spray the device again and wait for 5 minutes disinfectant contact time 8 Connect the Cube 25 NIV to Res...
Page 13: ...cordance with the Manufacturer s written instructions In the event of the warranty becoming void as a result of an occurrence specified in paragraph 3a to 3d above the Manufacturer will provide the Customer with a written notice of its reason for rejecting the warranty claim The Customer shall in accordance with the Supplier s written instructions deliver the defective item s at the Customer s exp...
Page 14: ...oc Ref ETION MED MDF 01 06 08 Rev02 Authors Contributors Date Doc Ref Jaco Botha and Sanet Groenewald Andre van Tonder Andre Ten Napel Björn Kleinschmidt Jaco Basson 2020 10 14 ETION MED MDF 01 06 08 Rev 02 ...