Generator G11
40
Generator G11 Operator’s Manual
Rated maximum output power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d
= 1�2
√
P
80 MHz to 800 MHz
d
= 1�2
√
P
800 MHz to 2.5 GHz
d
= 2�3
√
P
0�01
0�12
0�12
0�23
0�1
0�38
0�38
0�73
1�0
1�20
1�20
2�30
10
3�79
3�79
7�27
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer�
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies�
Note: These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people�
Go to
www.e-ifu.com
for latest revision of service manual�
Footswitch
Non-ionizing Radiation
Equipotential Grounding
Electrical and Electronic equipment� Return waste to a collection system or treatment and recycling facilities�
Applicable in the EU� Follow decontamination instructions before returning waste�
Authorized Representative in the European Community
Authorized Representative in the USA
Manufacturer / Date of Manufacture