? Can’t power on
Check whether the adaptor is well plugged and the 5VDC power supply is working.
ERROR CODE AND SOLUTIONS
CODE
PROBLEM
E01
E02
E03
E17
E31
.
34
CO
2
or PM2.5 sensor is damaged
Measured reading is under the lower limit
Measured reading is above the upper limit
ABC mode of CO
2
sensor is fail
Temp
./
RH measuring circuit is in error
Send back to repair
For CO
2
E02: Re-calibrate. For other E02: leave the meter in
room condition for 30 mins.
If it not works, send back to repair
For CO
2
E03: put the meter in fresh air for 5 mins. For other
E03: leave the meter in room condition for 30 mins.
If it not works, send back to repair
Send back to repair
Send back to repair
SOLUTION
5. Under the stable atmospheric environment, the sudden fluctuation of air humidity would have strong
impact to accumulative rise and abrupt rise of PM concentration.
The strong fluctuation in RH is an essential condition for 32 times higher of abrupt rise in PM
concentration so the probability of abrupt rise in PM2.5 concentration is as high as more than 80%
due to the aftereffect of strong RH fluctuation
Please note, extremely high (ex: 85%) or low RH (35%) could not have big effect on the change in PM
concentration but the fluctuation of RH can . This explains why the PM2.5 value abrupt rise while the
wheather is suddenly changed.
? Unexpected or expected blackout
While meter is programmed to record data, once the logging is started, blackout will cause the logging function
stop. “StoP” and real time CO
2
value will appear in turns.
Please return to computer to download data and re-program again.
? NO response at all
The meter might crash after long term usage. There are two methods to power on
again. One, unplug the power cable for 2 hours and then plug again. If you could
not wait for 2 hours time to discharge, please use any suitable and available pin
to do the hardware reset. The reset hole is on the rear side of meter, next to PM2.5
Air In filter. Using pin to press the reset key and see the display re-start again.
6.While see the need or after replacing filter, you can manually clean up the PM2.5 sensor periodically.
Press “
SET
” + “ ” keys at the same time to activate the sensor cleaning function for 10 seconds.
Please do it in clean environment.
/
II
WARRANTY & RETURN AUTHORIZATION
The meter is warranted to be free from defects in material and workmanship for a period of one year from the
date of purchase. This warranty covers normal operation and does not cover misuse, abuse, alteration,
neglect, improper maintenance, or damage resulting from leaking batteries. Proof of purchase is required for
warranty repairs. Warranty is void if the meter has been opened.
Authorization must be obtained from the supplier before returning items for any reason. When requiring a RA
(Return Authorization), please include data regarding the defective reason, the meters are to be returned along
with good packing to prevent any damage in delivery and insured against possible damage or loss.