-17-
state, then
continue measurement.
In the
MR
state, need to press
MEM
key to exit the
MR
state, then
continue to measurement.
In setting Alarm Critical Value state, need to press the
ON/OFF
key
or press the
AL
key for 3 seconds, exit Alarm Critical Value state,
then continue to measurement.
In the current test mode, press
Ω
key to switch to resistance test
mode.
4
.
Current Measurement
After the booting auto-inspection is completed, the Meter auto-
matically enter the resistance measurement mode. Upon showing
"
OL Ω
", press
A
key, and the Meter enter the current measurement
mode, showing "
AC 0.00mA
", see Figure 7. At this point, press the
trigger and open the jaws, clamp the target wire, reading to get the
current value.
It shows "
OL A
‖, indicating
that the measured current
value exceeded the upper
limit of Meter, see Figure 8.
Flashing display symbols , go with intermittent "beep
--beep--beep--" sound, indicating the measured current exceeds
the current of Alarm Critical Value.
In the
HOLD
state, need to press
HOLD
key to exit the
HOLD
state, then
continue measurement.
In the
MR
state, need to press
MEM
key to exit the
MR
state, then
continue to measurement.