ETCR 7100 Manual Download Page 10

 

-8-

3.    Data Access/ Exit 

In test mode, press 

PEAK

 and 

POWER

 key to access data inquiry form 

group 

R

01

,  and  display 

MR

  symbol.  Press 

PEAK

  or 

POWER

  key  to 

increase  or  decrease  the  page  number,  it  will  automatically  return  back  to 

group 01 when access the last group. 

Press 

HOLD

 key to exit date inquiry, back to test mode.

 

4.    Data Upload 

Connecting  the  meter  and  computer  with  USB-RS232  communication 

line  attached  in  package.  Start  up  the  meter,  run  software,  choose  history 

access, then read, save, report, print history data, etc. The more data storage, 

take the longer time to read it. Historical data can be saved in Txt text or Excel 

format. 

5.    Delete Data 

In the date inquiry mode, press 

PEAK

 and 

POWER

 key to delete all the 

stored  data,  and  return  back to test mode.  At the  same time, 

dEL

  symbol 

display. 

6.    General Measurement 

During  the  measurement,  the  LCD  real-time  display  the  value  of 

measured  current  and leakage  current, the  value  displayed on  LCD  change 

with different current and leakage current. No measured value reserved when 

take the meter away from measured wires, LCD display zero. 

 

 

High  voltage,  very  dangerous

Only  qualified  personnel 

after  training  could  conduct  operation  on it.  The operator 
should obey safety regulations; Otherwise there will be the 
danger  of  electric  shock  resulting  in  personal  injury  or 
casualty.

 

 

Dangerous!  Can  not  be  used  to  test  voltage  higher  than 
600V. Otherwise there will be the danger of electric shock 
resulting in personal injury or casualty.

 

Summary of Contents for 7100

Page 1: ......

Page 2: ... 3 III Model 4 IV Electrical Symbols 4 V Technical Specification 4 VI Instrument Structure 6 VII LCD Display 7 VIII Method of Operation 7 1 Switch On Off 7 2 Data Hold Storage 7 3 Data Access Exit 8 4 Data Upload 8 5 Delete Data 8 6 General Measurement 8 7 PEAK Measurement 10 IX Battery Replacement 10 XI Accessories 11 ...

Page 3: ...nts in wet places Do not take the measurements in the presence of explosive gas and combustibles or in dusty places Avoid any contact with the circuit under test even though you are not taking any measurement Avoid any contact with exposed metal parts unused measuring terminals circuits etc Do not take any measurement whenever anomalous conditions occur such as deformations breaks leakages blind d...

Page 4: ...replaced with identical ones Do not test nor connect to any circuit exceeding the specified overload protection Do not take measurements under environmental conditions exceeding the limits indicated in this manual Make sure that batteries are correctly installed 2 During Use You are recommended to read carefully the following instructions CAUTION Failure to comply with warnings and instructions ma...

Page 5: ...m 4mm flat cable and steel earth wires It is suitable for leakage current measurement of cable and transformer earth steel Full automatically and LCD display all the data are displayed in the screen which is very clear and convenient The meter is widely used in electric power communication meteorology railway oil field architecture measuring teaching research unit industrial mining enterprises etc...

Page 6: ...rules otherwise there is risk of electric shock resulting in bodily injury or fatalities Warning Safety rules must be strictly abided by otherwise personal injury or equipment damage may be caused Alternate Current AC Direct Current DC Double Insulation V Technical Specification Function Measure AC leakage current big current particularly suitable for leakage current measurement of cable and trans...

Page 7: ...shift Sample Rate About 2 times second Line Voltage Below AC 600V line measurement Display Mode 4 digital LCD display length 47mm width 28 5mm Meter Size Length 350mm Width 180mm Height 55mm Backlight Suitable for dim places Data Hold HOLD symbol appears Overflow OL symbol appears Automatic Shutdown Automatically shutdown about 5 minutes after power on to reduce battery consumption Voltage Detecti...

Page 8: ...AT III 600V VI Instrument Structure 1 Clamp 108mm 148mm 2 Lock switch after lock the clamp can t be open 3 HOLD key 4 PEAK key 5 LCD display 6 RS232 interface 7 Opening lever 8 POWER key 9 Battery cover 10 Battery cover screw 1 piece 11 Up and down cover connecting screws 6 pieces ...

Page 9: ...ode press POWER key to switch off The meter will automatically power off after booting 5 minutes later If LCD display is darker maybe the battery voltage is too low please replace batteries In data hold mode firstly press HOLD key to cancel the lock then press POWER key to switch off 2 Data Hold Storage In test mode press HOLD key to lock currently displayed value and display HOLD symbol At the sa...

Page 10: ...lete Data In the date inquiry mode press PEAK and POWER key to delete all the stored data and return back to test mode At the same time dEL symbol display 6 General Measurement During the measurement the LCD real time display the value of measured current and leakage current the value displayed on LCD change with different current and leakage current No measured value reserved when take the meter ...

Page 11: ...rly maintain the meter Clamp live wire and null line together to measure leakage current of electric equipment Note 2 wires Clamp earth wire to measure grounding line leakage current of electric equipment Note single wire Clamp four wires of there phase to measure the total leakage current Note 4 wires Clamp main line to measure total current of that main line Note single wire Test reference illus...

Page 12: ...D In general test mode press PEAK key to access or exit PEAK test mode symbol displays in PEAK measurement mode In other mode should first exit and return back to test mode and then operate PEAK measurement as above steps IX Battery Replacement Warning Make sure the battery cover is well closed before measurement otherwise there will be danger Take note of the battery polarity otherwise it may cau...

Page 13: ... are successfully replaced repeat step 2 if it doesn t work XI Accessories Main Unit 1 piece Meter Box 1 piece RS232 Data Line 1 piece Disk 1 piece Battery 4 pieces Alkaline Dry Battery LR6 User Manual 1 piece Guarantee Card 1 piece Certification 1 piece ...

Page 14: ... Electronic Technology Company Address F 3F No 4 Pengshang Zhifu Road Jiahe Baiyun District Guangzhou Guangdong China Post Code 510440 Tel 86 20 62199556 62199554 Fax 86 20 62199550 E mail info etcr cc Website www etcr cc ...

Reviews: