ENGLISH
15
INITIAL DISPLAY
NOTE FOR THE PROGRAMMER:
Read the manual before starting programming or always have it at hand to be sure you are making the
correct selections.
IMPORTANT: if no keys are pressed for 60 seconds, the instrument will show the current measurement.
To go forward quickly,
press and hold
one of the
◄►
buttons
The software version is shown when the instrument is on the lower part of the display.
The software is subject to revisions without notice.
The instrument is prepared for measuring and is then ready to operate.
At this stage some status messages might be displayed such as:
S1 S2 L1 DATE/TIME AUX OVERDOSE ALARM MAX DOSE FLOW BASIC MENU EXPERT MENU
which might be active due to the current measurements, just go Forward with programming.
The
INITIAL DISPLAY
shows the chlorine Cl ppm measurements according to the range selected in the
paragraph.
SELECT THE CHLORINE RANGE”
. If the instrument has already been programmed, the display
shows the programs selected previously.
The
CONTINUOUS MEASUREMENT DISPLAY
shows the measurements of the parameter, the status of the
functions and the alarm indications.
The
MAIN DISPLAY
shows the measurements of the parameter, the status of the functions and the alarm
indications.
When the instrument is
switched on for the first time
, a list of functions of all keys is displayed. This screen
is no longer displayed during subsequent start-up operations.
READ THE MESSAGE CAREFULLY, THEN PRESS
O K
K
TO START.
USE OF THE KEYS
OK: access to the menu, start/stop, selection and editing
selects the step of the menu or increases / decreases the value
ESC: goes back to the previous menu or does not save the change
* Press OK to start*
To go forward quickly,
press
and hold one of the ◄► buttons
IN ENGLISH
S el ect the l anguage of th e P R O G R A MM I N G M E N U
To select the language of the programming menu.
After selecting the Language (ITALIAN, ENGLISH), the programming menu
adapts accordingly.
S el ect the C H LO R I N E M EA S U R E M EN T R A NG E
The instrument of the
eSelect M CL
series offers the option of selecting 4 chlorine measurement ranges and related sensors:
Ion-selective membrane sensors with measurement ranges 0-2 Cl ppm (preset); 0-20 Cl ppm; 0-20 Cl ppm (Total chlorine);
Open type amperometric cell (model CLC) range 0-10 Cl ppm
After making the selection, the measurement parameters displayed will change accordingly.
ETATRON
Model eSelec M1
R e v . X . X
CONTROLLER SETTINGS
Language
Eng lis h
BASIC MENU
SETTINGS
Lingua
Ital iano
MENU BASE
Summary of Contents for eSelect M 1 CI