ETAS
Troubleshooting
AC Voltage Distributor ES4700.1-B
37
11
Troubleshooting
shows possible faults for the ES4700.1-B. When it is not possible to eliminate the
error cause, please contact ETAS (for ETAS contact information, refer to chapter 122 "ETAS
Fault
Error cause
Solution
Battery supply does not
work
Button "EMERGENCY
STOP" is pressed
Release the button
"EMERGENCY STOP"
THERMO connector is not
closed
Close the THERMO
connector
System is not running
Emergency button is
pressed
Release the button
"EMERGENCY STOP"
Main circuit fuse released
Enable fuse
RCD released
Enable RCD
Blown fuse "F7"
Replace fuse (on rear side)
DC +48V not available
Blown fuse "F10"
Replace fuse (on rear side)
DC +12V not available
Blown fuse "F11"
Replace fuse (on rear side)
DC -12V not available
Blown fuse "F12"
Replace fuse (on rear side)
DC +5V not available
Blown fuse "F13"
Replace fuse (on rear side)
FAN DC +12V not available
Blown fuse "F14"
Replace fuse (on rear side)
Tab. 11-1
Fault, error cause, solution