background image

3

Molift MRS Quattro / www.etac.com 

Product label

The Product label contain article number and 

production date.

MRS Quattro

SWL 300kg / 661lbs

Warnng labels and Symbols

Symbols used on the product, explained in more 

detail:

CE marked

Refer to user manual

Indoor Use only

Manufacturer

Date of manufacture

Catalogue number

Serial number

Medical Device

YYYY-MM-DD

Product identification

General

CE declaration

The product, and its related accessories, described in this instruction for use (IFU), is in compliance 
with the regulation (EU) 2017/745 of 5. April 2017 – as a medical device, risk class I.
Any serious incident that occurs in relation to the device should be reported to the manufacturer 
and the competent authority of the Member State in which the user and/or patient is established.

Conditions for Use

Hoisting and transferring a person may contain an element of risk. A risk assessment should be carried 

out prior to use and only trained personnel should use the equipment  as described in the user manual.

Modifications and use of components made by other manufacturers.

We recommend only using Molift components and spare parts. Declaration of conformity is not valid 

and Etac is not responsible for warranty if any modifications are made to the product. Etac shall not be 

liable for faults or accidents that can occur when using components made by other manufacturers.

Only certified personnel are allowed to perform repairs.

Only suitable trained people should install this equipment. 

The work contains handling of heavy loads.

Warranty notice

Two-year warranty against defects in workmanship and materials of our products. 

Please refer to www.etac.com for terms and conditions.

Summary of Contents for Molift MRS Quattro

Page 1: ...MRS Quattro EN Installation manual BM22001 Rev B 2021 06 03...

Page 2: ...11 Adjustable foot 12 In Rail Charging System 12 Maintenance 13 Cleaning and disinfection 13 Reconditioning 13 Recycling 13 Spare parts 13 Periodic Inspection 13 Periodic Inspection Scope 13 Troublesh...

Page 3: ...ght materials Traverse ranges from 2 to 3 5 meters and the lifting capacity is 300 kg 661 lbs It is ideal for hospitals and institutions where permanent installations are not possible or desirable but...

Page 4: ...ich the user and or patient is established Conditions for Use Hoisting and transferring a person may contain an element of risk A risk assessment should be carried out prior to use and only trained pe...

Page 5: ...6 0 lbs Total weight Complete unit 110 95kg 3x3m 131 55kg 4x4m Available boom lengths A 3m or 4m Dimensions Width D 3 16 4 16m Height B 2 33m 2 33 2 38m with adjustable foot Depth E 3 16 4 16 m Max i...

Page 6: ...the traverse boom 4 Tighten the bolts loosely Mounting Quattro Frame Tip To facilitate the installation make 4 support blocks approximately 150 mm high 1 Make sure you have free space of aprox 3x4 met...

Page 7: ...both ways 2 If not loosen the bolts and adjust 3 When the readings is the same the frame is square or rectangle 4 Tighten all bolts and nuts Warning for lifting of heavy loads Dropping the frame can c...

Page 8: ...and attach legs at the opposite corner Mounting of legs Ceiling height 2 4 m Start from step Perpendicular frame 1 Detach one boom and keep the corners on 2 Move the traverse boom to a position close...

Page 9: ...nuts 11 Release the traverse boom Check list before use Don t start using MRS Quattro before completing control accord ing to checklist after assembly Use this checklist to verify that the MRS Quattro...

Page 10: ...9 Molift MRS Quattro www etac com 1 Attach the adjustable feet 2 Adjust the levelling feet for horizontal position of MRS Quattro Unscrewing the foot too much can cause the foot to come loose Assembly...

Page 11: ...efore lifting patient Never lift the user higher than necessary to carry out a lift Never leave a user unattended in a lifting situation Transport and storage MRS Quattro can be stored and transported...

Page 12: ...WL Molift Air 205 Molift Air 300 Molift Air 205 Propulsion Molift Air 300 Propulsion Trolley for Molift Air MRS Trolley Basic Art no 2520000 MRS Trolley Propulsion Art no 2500103 Molift Nomad Molift N...

Page 13: ...s Adjustable foot Adjustable feet are available for MRS Quattro Adjustable foot for MRS Quattro Art no 1610200 In Rail Charging System Possible to use the in rail charging system with the MRS Quattro...

Page 14: ...immediately be taken out of service and marked clearly with out of order and shall not be used until the MRS Quattro is repaired Periodic Inspection shall be performed at least once a year or more fr...

Page 15: ...rly thightened Adjustable foot adjusted up too much Wobbly between Traverse boom and frame Check bolts between bracket and boom Make sure bolts are properly thightened Control bolts between bracket an...

Page 16: ...Etac A S Parallelvej 3 DK 8751 Gedved www etac com...