21/70
This function proves to be particularly useful when cleaning large-
sized parts.
Press the Sweep button to turn on and off this function.
The Degas and Sweep functions cannot be activated simultaneously.
7.3 Post-cleaning
Once the cleaning step has been completed, rinse all parts under
running water.
Drain the tank whenever the solution becomes dirty or if the machine
has not been used for a long period of time, as certain types of
residue can indeed damage the tank.
8. Cleaning
products
Select a cleaning product compatible with the ultrasonic solution.
Without ensuring such compatibility, the tank could be damaged and,
under a worst case scenario, the user could be injured.
Be sure to heed all restrictions concerning the use of solvents.
Should any doubt arise, contact your distributor.
Any damage due to a failure to respect the restrictions cited in
Chapter 8.2
will not invoke the manufacturer's liability and therefore
will not be covered by the product warranty.
8.1
Restrictions concerning the use of solvents
Caution! Never pour, under any circumstances, inflammable
liquids or solvents directly into the tank due to the fire and
explosion risk!
The presence of ultrasound stimulates the liquid evaporation process,
which serves to form a light mist above the liquid surface, leading to a
risk of ignition should the device be positioned near a heat source.
It is strictly
prohibited
to insert any of the explosive and inflammable
substances listed below into the tank:
•
products branded, in accordance with EC directives, with a
symbol or safety precaution R1 - R9, or any of the E, F+, F, O or
R10, R11 or R12 branding for inflammable substances.
Handling of an inflammable liquid is allowed (up to a 1-liter maximum)
provided all general safety precautions are strictly adhered to and the
following directives respected:
- Operations are to be carried out in a well-ventilated zone, within a
separate container (e.g. glass beaker), itself placed into the tank filled
with a non-inflammable liquid (water + detergent).
Procedure tip
Post-cleaning
Drainage
Liability
Exception