ESSILOR UMCP550 V1 OCT2015
19
After an automatic stop, the device returns to normal mode.
12. Storage and maintenance
12.1. Handling the device
When the projector is not in use, protect it with the cover provided by the manufacturer.
Do not touch any of the optical parts with your fingers and avoid exposing any parts to dust. The dust accumulated on
any optical part of the device is to be removed using a brush or a bellows. Should you notice a finger smear or trace
of grease on an optical part, apply an alcohol-based cleanser and wipe with a lens cloth or gauze.
If the projector surfaces are dirty, take the precaution of cleaning them with a dry cloth. For marks that are
difficult to remove, we recommend using a damp cloth or neutral detergent.
Avoid employing organic solvents as these would damage the paint coating of the device.
If the projector remains idle for awhile, unplug the power supply cable.
Also, if the projector remains idle for awhile, remove the batteries from the remote control.
12.2. Replacing a fuse
3- Unplug the power supply cable
(8)
to avoid any risk of electrical discharge.
4- Remove the fuse holder
(9)
, then replace the fuse.
If you are not using the appropriate fuse, the device will tend to easily malfunction or potentially cause an
accident. Only use the prescribed fuse (250 V, 2 AT).
The fuse holder contains two fuses. Since you cannot visually identify the defective fuse, remove both and test
them individually.
Anticipate other problem sources should the device continue to malfunction following replacement of the fuse.
Immediately unplug the power supply cable and contact the closest product retailer.
12.3. Replacing batteries on the remote control
Pull on the battery compartment lid, then remove the used batteries and replace them with new ones. Position the
new batteries as indicated on the sticker at the bottom of the battery compartment. Use two non-corrosive batteries
in order to avoid leaks (x 2 UM-4 alkaline batteries).
13. Trouble-shooting
Be sure to check each of the following elements before bringing the device in for repairs. If, once these elements
have been verified, the instrument is still not functioning properly, contact your product retailer.
4- The lamp does not light up when the main switch
(7)
is turned on.
Cause 1: Poor connection of the power supply cable
Place the switch on the "0" position and then firmly plug in the power supply cable
(8)
.
Cause 2: The fuse has blown.
Replace the fuses (see
5.3 – Replacing a fuse
).
Summary of Contents for CP550 Series
Page 2: ...Ref UMCP550 V1 OCT2015...
Page 12: ...ESSILOR UMCP550 V1 OCT2015 10...
Page 13: ...ESSILOR UMCP550 V1 OCT2015 11 TEST PROJECTOR USER S MANUAL...
Page 14: ...ESSILOR UMCP550 V1 OCT2015 12...
Page 24: ...ESSILOR UMCP550 V1 OCT2015 22...
Page 25: ...ESSILOR UMCP550 V1 OCT2015 23 SEHZEICHENPROJEKTOR BEDIENUNGSANLEITUNG...
Page 26: ...ESSILOR UMCP550 V1 OCT2015 24...
Page 36: ...ESSILOR UMCP550 V1 OCT2015 34...
Page 37: ...ESSILOR UMCP550 V1 OCT2015 35 PROYECTOR DE TESTS MANUAL DE USO...
Page 38: ...ESSILOR UMCP550 V1 OCT2015 36...
Page 48: ...ESSILOR UMCP550 V1 OCT2015 46...
Page 49: ...ESSILOR UMCP550 V1 OCT2015 47 PROIETTORE PER TEST MANUALE D USO...
Page 50: ...ESSILOR UMCP550 V1 OCT2015 48...
Page 60: ...ESSILOR UMCP550 V1 OCT2015 58...
Page 61: ...ESSILOR UMCP550 V1 OCT2015 59 PROJECTOR DE TESTES MANUAL DE UTILIZA O...
Page 62: ...ESSILOR UMCP550 V1 OCT2015 60...
Page 72: ...ESSILOR International 147 avenue de Paris 94220 Charenton le pont France...
Page 73: ......
Page 75: ...Ref UMCP550 V1 OCT2015...
Page 85: ...ESSILOR UMCP550 V1 OCT2015 10...
Page 86: ...ESSILOR UMCP550 V1 OCT2015 11 TEST PROJECTOR USER S MANUAL...
Page 87: ...ESSILOR UMCP550 V1 OCT2015 12...
Page 97: ...ESSILOR UMCP550 V1 OCT2015 22...
Page 98: ...ESSILOR UMCP550 V1 OCT2015 23 SEHZEICHENPROJEKTOR BEDIENUNGSANLEITUNG...
Page 99: ...ESSILOR UMCP550 V1 OCT2015 24...
Page 109: ...ESSILOR UMCP550 V1 OCT2015 34...
Page 110: ...ESSILOR UMCP550 V1 OCT2015 35 PROYECTOR DE TESTS MANUAL DE USO...
Page 111: ...ESSILOR UMCP550 V1 OCT2015 36...
Page 121: ...ESSILOR UMCP550 V1 OCT2015 46...
Page 122: ...ESSILOR UMCP550 V1 OCT2015 47 PROIETTORE PER TEST MANUALE D USO...
Page 123: ...ESSILOR UMCP550 V1 OCT2015 48...
Page 133: ...ESSILOR UMCP550 V1 OCT2015 58...
Page 134: ...ESSILOR UMCP550 V1 OCT2015 59 PROJECTOR DE TESTES MANUAL DE UTILIZA O...
Page 135: ...ESSILOR UMCP550 V1 OCT2015 60...
Page 145: ...ESSILOR International 147 avenue de Paris 94220 Charenton le pont France...
Page 146: ......