Care@Home™ 

Emergency Pendant Advanced 

User Guide 

ESUGSC032 

Version 1.4 

December 2017 

Summary of Contents for CareHome ES700EPA

Page 1: ...Care Home Emergency Pendant Advanced User Guide ESUGSC032 Version 1 4 December 2017...

Page 2: ...ing with the Active Service 5 3 1 Active Service Settings 8 3 2 Outgoing Call Settings 8 3 3 Changing the Emergency Call Number 9 4 Testing Fall Detection 10 5 Operating the EPA 11 5 1 Notifications 1...

Page 3: ...supplied lanyard When the EPA detects a fall or the EPA button is pressed the EPA sends an emergency message via a radio signal to the Control Panel CP which in turn passes the message to the monitori...

Page 4: ...g the EPA move about 50 cm 20 in away from the CP 3 Press the EPA button 4 Ensure that the pairing process is successful The process can have the following results Success The CP beeps and the EMERGEN...

Page 5: ...resident s mobile device meets the following requirements Apple iOS 9 0 or later Android OS 4 or later Minimum mobile screen size of 4 7 To use the Care Home Active Service you must install the Care...

Page 6: ...Installing with the Active Service 6 Care Home Emergency Pendant Advanced User Guide The following shows the app in operation...

Page 7: ...Installing with the Active Service Care Home Emergency Pendant Advanced User Guide 7 Figure 3 Pairing with the Active App...

Page 8: ...ttings If you tap Settings in the app the following is displayed Figure 4 Active App Settings Caution Do not change the DNS settings 3 2 Outgoing Call Settings You can change the outgoing call setting...

Page 9: ...to the telephone number specified by your service provider Off the emergency alarm call goes to the telephone number specified at Emergency Call Number 3 3 Changing the Emergency Call Number To chang...

Page 10: ...endant Advanced User Guide 4 Testing Fall Detection Testing the EPA is optional To test the EPA 1 While sitting hit the EPA on your thigh three times 2 Put the EPA down flat on a table top Wait 25 sec...

Page 11: ...CP is configured the EPA periodically sends information to the CP including The temperature The EPA battery charge percentage 5 2 LED Indications The EPA LED lights up as shown in the following table...

Page 12: ...The EPA is a supervised device The EPA periodically transmits its status to the CP The frequency of the reporting periods can be configured remotely If after a pre defined time the CP does not receiv...

Page 13: ...the Battery The EPA uses a 3 V CR2450 lithium battery To replace the battery 1 Remove the lanyard 2 Open the EPA by inserting and twisting a coin to remove the cover Figure 6 Open the EPA Caution Avo...

Page 14: ...5 Ensure the top indicator on the EPA is aligned with the top indicator on the lanyard Caution The EPA must be properly aligned for its fall detection capability to function properly 6 Insert the new...

Page 15: ...ence environment Encoding 32 bit ID over 4 billion combinations Power 1 x 3 V CR2450 lithium battery Up to two year battery life NOTE Nominal battery life is computed according to Essence s standard u...

Page 16: ...C has similar functionality and specifications to the EPA The only difference is in how the EPA BC communicates with the CP when it detects a fall or the panic button being pressed However just as wit...

Page 17: ...functionality to the EP but has the same replaceable battery and physical characteristics as the EPA The EP Plus does not include fall detection and can be worn as a pendant or on the wrist The LED in...

Page 18: ...pply to you some or all of the above disclaimers exclusions or limitations may not apply to you By using the information contained herein you agree that the laws of the State of Israel without regard...