Criterion Series Benchtop User’s Manual
31
5.4. Specimen Safety Device Setup
Overheat Protector
CAUTION
•
Always set the overheat protector BEFORE
beginning tests. Set it according to the type of
specimens and test conditions.
Unless properly set, the overheat protector will not trip in the
event of cabinet trouble, which could lead to specimen damage.
Procedure:
Turn the dials until a suitable temperature is displayed.
OVERHEAT
PROTECTOR
Dial
Fig. 5.5 Overheat protector
Absolute High/Low Limit Temperature (Humidity) Alarms
CAUTION
•
Always set the absolute high/low limit temperature
(humidity) alarms BEFORE beginning tests. Set
alarms according to the type of specimens and test
conditions.
Unless properly set, the absolute high/low limit temperature
(humidity) alarms will not be triggered in the event of cabinet
trouble, which could lead to specimen damage.
Summary of Contents for BTL-433
Page 1: ...CRITERION CHAMBER SERIES BENCH TOP INSTRUCTION MANUAL...
Page 17: ...Criterion Series Benchtop User s Manual 14...
Page 23: ...Criterion Series Benchtop User s Manual 20...
Page 31: ...Criterion Series Benchtop User s Manual 28...
Page 41: ...Criterion Series Benchtop User s Manual 38...
Page 48: ...Criterion Series Benchtop User s Manual 45 8 Appendix 8 1 Specifications...