Precautions
With the settings in Table 1.2., the safety devices will trip in the succession shown in
Figure 1.2. as temperature inside the chamber rises beyond the target temperature. The
first to trip will be the upper deviation limit alarm, followed by the absolute high limit alarm
and then the overheat protector. Due to the fact that the chamber is equipped with
multiple safety devices, abnormal temperatures cannot escape detection even if one of the
devices is not working properly.
80
60
75
70
Chamber Temperature (°C)
Time
Abnormal te mp erature
detected here. (If 1
is not working)
Abnormal
temperatu re
detected h ere.
Alarm cleared here
3. Ov erheat protector
2. Absolute hig h limit
tempeature alar m
1. Upper dev iation limit
temperature alarm
Target tempera ture
Abnormal temperatu re detected here
(If 1 and 2 are not working)
Fig. 1.2 Temperature Alarm Trigger Conditions