Field test method:
Test Schematic
During on-site testing, the test panel of the instrument should be placed where the ESD
sensitive components are actually operated.
The specific test method is subject to the instructions of different test instruments
.
09
Static Elimination Performance
Operating Voltage: DC24V
Test Voltage ±1000V-±100V
Positive
Negative
Positive
Negative
Distance(mm)
Static elimination
time
Residual voltage
Measurement method: use 20PF, 150mm static elimination monitor to measure.
When the air volume is maximum, the dissipation time from ±1000V---±100V.
Different environment test results are different, the above is the reference data, please test
according to the actual use environment
.
±20V
300
2.5S
3.0S
600
7.5S
8.9S
900
15.5S
17.6S
Temperature23±3
O
C
Humidity50%±5%
After the discharge emitter is used for a long time, there will be dust and other pollutants in and
around the discharge emitter.
Use cotton swabs to remove contaminants.
The discharge emitter should be cleaned regularly (approximately 100 hours of use need to
be cleaned once). When the pollution is serious, the cotton swab can be cleaned with alcohol.
Be careful not to damage the discharge emitter when cleaning.
Cleaning of the discharge emitter
10
Routine Maintenance