10
3.4.2 For Unit with Ducting
For ducted Unit, you will need to cut a portion of the bag to accommodate the ducting system. The
decontamination method itself is basically the same as the non-ducted Unit shown above with some additional
steps, namely:
Close the air tight damper
Wrap the bag around the ducting system and seal it using aluminum tape. Use multiple layer of
aluminum tape to make sure that it is sealed correctly.
Create a rounded shape using aluminum tape so the adhesive part is on the outside. Use it to put the
bag together.
Seal the connection with multiple layers of aluminum tape to ensure its leaktightness.
The BSC is ready for formalin
decontamination.