5
Formalin Vaporizer
2.2.1.7 Reset Default
This will start the decontamination process. The process will start with pre-processing step, and continue with
formalin vaporizing step, contact time step, ammonia vaporizing step and finally neutral time step.
RESET DEFAULT
MENU
READ MANUAL | PRESS SET
2.2.2 After Decontamination Process Start
This menu can only be access during the decontamination process.
2.2.2.1 Resume Step
This menu allows you to resume the decontamination process after a pause i.e. by accessing menu.
If there’s a power failure during the decontamination process, the formalin vaporizer will automatically
resume the decontamination process once the power is recovered. In this case, the operator doesn’t even
need to access this menu.
RESUME STEP
MENU
CONFIRM?
2.2.2.2 Skip Step
This allows you to skip a step in the decontamination process.
SKIP STEP
MENU
SKIP STEP?
2.2.2.3 Cancel Process
This allow you to cancel the decontamination process altogether.
CANCEL STEP
MENU
CANCEL PROCESS?