MIRI
®
TL family`s multiroom IVF incubators User Manual Rev. 1.0
77
24 Temperature validation
The MIRI
®
TL6 multiroom IVF incubator is equipped with 6 PT-1000 Class B sensors,
whereas the MIRI
®
TL12 multiroom IVF incubator – 12,
that are located in the center of
the bottom of each compartment.
Figure 24.1
PT-1000 Class B sensors in the MIRI
®
TL6 multiroom IVF incubator
Figure 24.2
PT-1000 Class B sensors in the MIRI
®
TL12 multiroom IVF incubator
These sensors serve external validation purposes. They are completely separated from
the circuit of the unit.
Temperature conditions in the compartments can be continuously logged through the
external connectors on the unit's side without compromising its performance.
Any logging system that uses standard PT-1000 sensors may be used.
Esco Medical can supply an external logging system (MIRI
®
-GA12) for the sensors.
25 Gas level validation
Gas concentration in each compartment of the MIRI
®
TL6 or MIRI
®
TL12 multiroom IVF