and the chambers. Over time, the lines' performance can be affected, causing more extended
recovery time due to clogging.
Therefore, the gas lines must be replaced once every 3 years to maintain the fast recovery
time after lid opening.
Please follow these safety precautions when changing gas lines:
•
Always use original gas lines (contact Esco Medical or your local distributor for more
details or ordering).
•
Change the gas lines within 3 years from the date of installation.
•
Failure to change the gas lines may cause slow recovery times or breakdowns.
•
Warranty void if wrong/ non-original gas lines are used.
Please refer to the service manual for replacement instructions.
37.11 Flow sensors
The flow sensors are used by the CO
2
/N
2
regulations and for logging the unit's gas
consumption.
This sensor's lifetime is more than 3 years, but Esco Medical recommends the sensor to be
replaced once every 2 years for safety reasons.
Please follow these safety precautions when changing sensors:
•
Always use an original flow sensor (contact Esco Medical or your local distributor for
more details or ordering).
•
Change flow sensors within 2 years from the date of installation.
•
Failure to change the flow sensors on time may result in low/no CO
2
and O
2
gas con-
centration regulation.
•
Warranty void if wrong/ non-original sensors are used.
Please refer to the service manual for replacement instructions.
37.12 Pressure regulators
The internal pressure regulators protect the system from too high external gas pressures
that would damage the gas circuit's sensitive parts. If the pressure regulators are worn, they
may begin to drift and not offer the protection they are supposed to. It could cause
breakdowns or leaks in the internal gas circuit. Therefore, the regulators must be replaced
once every 4
years to maintain the system safe and stable.