eschmann TD830 Instructions For Use Manual Download Page 22

P22/37

E-IM54f

5.55 The output control settings required for a particular
surgical procedure are also dependent on a number of
external factors (type of tissue, speed at which the surgeon
works, type of electrode in use, area of forcep tips in
contact).  Therefore the recommended settings given in
Table 1 are for guidance only.

Alarms

Alarm Conditions

5.56 The audiovisual alarm, two-tone with   flashing
displays at one second intervals, will operate if:

  ♦

The non-divided plate electrode is not connected via
the plate cable to the unit, or the plate electrode and/
or cable is faulty (PCM).

  ♦

A divided plate cable is plugged into the unit and the
divided plate electrode is not making a sufficiently
good contact with the patient’s body (PAM).

  ♦

A low impedance path exists between the patient’s
body and earth (PEM). This condition can be caused
by the presence of an earthed ECG electrode, or by
an accidental contact between the patient’s body and
a conductive object such as part of the operation table
or a drip stand.

WARNING

The patient earth monitor (PEM) will not detect
a low impedence path between the patients
body and earth when the electrosurgical output
is activated.

  ♦

Excess power is detected (EPM) or dangerous
voltages are detected (PVM).

  ♦

More than one footswitch or fingerswitch is actuated
for monopolar USER 1, USER 2 or bipolar, e.g.

  ♦

fingerswitch and footswitch both actuated on the
monopolar USER 1 output.

  ♦

or, two fingerswitches on monopolar USER 2.

  ♦

or, pneumatic and electric footswitch on bipolar
output.

  ♦

or, any other forbidden combination of footswitch
and fingerswitch commands.

  ♦

Internal unit errors are detected (!)

In any alarm mode, all outputs are inhibited and alarm
volume is maximum, regardless of volume control setting.
If the unit is in standby mode only the bipolar display will
flash.

5.57 If more than one output demand is received at the
same time (such as two footswitches or a footswitch and a
fingerswitch button being pressed) the unit will produce
an alarm indication and no power output.  This is an
important safety feature and applies to any combination of
footswitch and fingerswitch for monopolar USER 1, or for
USER 2 or for bipolar.

Alarm identifications and remedies

5.58 Alarm identifications are displayed on the front panel
(see Fig. 11 A1-A7) and described in section 3, remedies
are as follows:

If the PCM or PAM circuit has been activated.

Remedy:

(i)

Check plate electrode is connected to unit.

(ii) Check plate electrode and cable for condition, if

faulty, substitute new plate electrode and/or cable.
This procedure should normally be carried out before
plate electrode is attached to patient.

(iii) If a divided plate electrode is being used check that

the divided plate electrode is making good contact
with the patient’s body.

If the PEM circuit has been activated.

Remedy:

(i)

Check for accidental direct contact between patient
and an earthed object, (drip stand or conductive part
of operation table).

(ii) If applicable disconnect suspect ECG electrodes.

Note:  An activated PEM alarm will automatically
be reset by releasing a depressed footswitch or
fingerswitch.

If the 

 circuit has been activated

Remedy:

(i)

The alarm will operate if the bipolar pneumatic
footswitch and the bipolar electrical footswitch are
pressed simultaneously. The alarm will also sound
if both the cut and coag footswitches, or if one of
the two footswitches and a fingerswitch button are
pressed simultaneously for USER 1, or if both
fingerswitch buttons are pressed for USER 1 or
USER 2.  The circuit will automatically reset when
ALL activations stop from ALL switches. This is an
important safety feature and applies to any
combination of footswitch and fingerswitch.

Summary of Contents for TD830

Page 1: ...TD830 ELECTROSURGICAL UNIT Instructions for use E IM54f 698257...

Page 2: ...Sussex BN15 8TJ England Tel 44 0 1903 765040 Fax 44 0 1903 875711 Overseas Customers Contact your local distributor In case of doubt contact Eschmann Equipment Patents and Trade marks The ESCHMANN nam...

Page 3: ...ns 14 Electrosurgical interference 14 Using two electrosurgical units 15 Using a flexible endoscope 15 Servicing 15 References 15 5 OPERATION Routine check 16 Connection of accessories 16 Monopolar 16...

Page 4: ...is most important that fuses of the correct type size and rating are installed see Technical Data CAUTION Read these Instructions for use carefully before using this electrosurgical unit and note ALL...

Page 5: ...25V min T10A Fuse type 20 mm Bipolar Carrier frequency 785kHz nominal square wave Power control Variable pulse group modu lation set by front panel control Amplitude set by micro macro range buttons L...

Page 6: ...rd edition General classification Class 1 Type CF Defibrillator proof Drip proof IPX 1 Patient leakage risk current always less than 100 microamps to earth ground from all patient circuits at 230V 50H...

Page 7: ...he symbol on the rear panel serial plate warns the user to read the accompanying documents the Instructions for use The symbol above sockets 25 26 27 and 28 denotes connection socket for a two button...

Page 8: ...Plate voltage monitor alarm has been activated EPM Indicates the Excess power monitor alarm has been activated Indicates the Multiple activation alarm has been activated Indicates an internal fault ha...

Page 9: ...heplateelectrodeisreliably attached to the patient s body and as close to the operating field as possible Store temporarily unused active electrodes such that they are isolated from the patient e g us...

Page 10: ...f the delivered power exceeds the allowed value by more than a certain margin Patient Earth Monitor PEM 3 5 The PEM circuit detects a low impedance path between the patient and earth It also reduces t...

Page 11: ...f the bipolar forceps and the tissue held between them is directly in view of the surgeon at all times 4 7 A plate electrode is not required and safety is increased even further as two power range out...

Page 12: ...ined throughout the surgical procedure The following points should also be noted Do not apply any plate electrode over bony protuberances metal implants folds of skin scar tissue hairy areas any form...

Page 13: ...ited number of re sterilization cycles Regular testing and inspection is therefore essential Do not sterilize and reuse single use types 4 32 Do not connect footswitches to the USER 1 output of the el...

Page 14: ...te electrode are kept a few inches away from the pacemaker pulse generator and its electrode leads while to avoid burns to the myocardium at the electrode site or possible ventricular fibrillation the...

Page 15: ...the following notes should be considered when writing a procedure for the use of this electrosurgical unit with an endoscope a To avoid the risk of potential burns caused by h f voltages on the body o...

Page 16: ...a supply of new plate electrodes is readily available Plate electrode With reusable types examine the platecableandplugforsignsofwear Ensurethatthe insulationoftheconnectingcableisingoodcondition Acce...

Page 17: ...olar electric socket 37 or a white pneumatic footswitch REF 83 581 68 Fig 10 to the bipolar pneumatic socket 36 Do not connect both types of footswitch at the same time 5 9 If the electrosurgical unit...

Page 18: ...uld be checked regularly in accordance with the Service Manual see section 1 8 OPERATING INSTRUCTIONS Note For numbers in brackets refer to Illustration 11 which folds out at the end of this manual Un...

Page 19: ...obtained with the normal power setting check the following before increasing power The plate electrode is correctly positioned and is in contact with the patient Theplateelectrodecableisintact andispl...

Page 20: ...non contact coagulation of tissue that tends to adhere to active electrodes Note Spray coag indicated with the bright green LED is not recommended for laparoscopic and endoscopic procedures where its...

Page 21: ...de or press macro bipolar selector touch button 5 to select macro bipolar mode The green indicators 2 or 4 will illuminate to show which bipolar mode has been selected 5 49 Set power output control 1...

Page 22: ...nit is in standby mode only the bipolar display will flash 5 57 If more than one output demand is received at the same time such as two footswitches or a footswitch and a fingerswitch button being pre...

Page 23: ...s been activated Remedy i If unit PVM alarm activates check for accidental direct contact between the active electrode and an earthed object ii Check that the plate electrode is correctly attached to...

Page 24: ...ation 40 to 60 TCRE TCAE Blend Normal endometrium 55 to 70 Transcervical Resection of Specialist cut Spongio endometrium 40 to 60 the Endometrium Ablation Blend Suspect fibroid inv 65 to 80 Spray coag...

Page 25: ...Power output watts Load resistance 200 ohms MONOPOLAR CUT POWER v CONTROL 0 50 100 150 200 250 300 0 25 50 75 100 125 150 175 200 225 250 Output control setting typical pow er Power output watts Load...

Page 26: ...0 ohms MONOPOLAR BLEND POWER v CONTROL 0 40 80 120 160 200 0 20 40 60 80 100 120 140 160 180 200 Output control setting typical power Power output watts Load resistance 400 ohms MONOPOLAR BLEND DIAGRA...

Page 27: ...ance 200 ohms MONOPOLAR SPECIALIST CUT POWER v CONTROL 0 50 100 150 200 250 300 350 0 30 60 90 120 150 180 210 240 270 300 Output control setting typical pow er Power output watts Load resistance 200...

Page 28: ...MONOPOLAR PINPOINT COAG POWER v CONTROL 0 20 40 60 80 100 120 0 10 20 30 40 50 60 70 80 90 100 Output control setting typical pow er Power output watts Load resistance 400 ohms MONOPOLAR PINPOINT COA...

Page 29: ...alue up to 10 Peak output voltage volts Open circuit MONOPOLAR SPRAY COAG VOLTAGE v CONTROL 0 500 1000 1500 2000 2500 3000 3500 4000 0 7 5 15 22 5 30 37 5 45 52 5 60 67 5 75 Output control setting typ...

Page 30: ...tput voltage volts Open circuit BIPOLAR MACRO POWER v CONTROL 0 10 20 30 40 50 0 1 2 3 4 5 6 7 8 9 10 Output control setting numerical value up to 10 Power output watts Load resistance 100 ohms BIPOLA...

Page 31: ...pow er Peak output voltage volts Open circuit BIPOLAR MICRO POWER v CONTROL 0 4 8 12 16 0 1 5 3 4 5 6 7 5 9 10 5 12 13 5 15 Output control setting typical pow er Power output watts Load resistance 10...

Page 32: ...Eschmann trained personnel otherwise the warranty for the equipment could be infringed 6 5 If the unit is to be stored for any length of time it should be plugged into the mains and switched on to ch...

Page 33: ...sposable 8 83 588 85 4 2mm pack of 5 8 REF Description Illustration No Colposcopy electrodes LLETZ cone biopsy 83 583 51 large loop 12mm pack of 5 8 83 584 19 large loop 15mm pack of 5 8 83 584 27 lar...

Page 34: ...34 37 E IM54f 83 122 42 83 122 46 83 122 49 83 207 56 83 122 93 83 109 39 83 109 47 83 581 09 83 581 13 83 581 17 83 581 25 83 140 20 4mm 8mm 83 141 01 83 141 36 83 141 28 83 141 44 1 2 3 4 5 83 109 7...

Page 35: ...8 m m 6 7 8 83 582 14 83 582 27 83 582 30 83 582 49 83 588 69 83 585 04 Disposable 83 585 11 Disposable 83 582 22 83 583 11 83 585 23 Disposable 83 588 77 83 588 80 83 583 51 83 584 19 83 584 27 83 5...

Page 36: ...P36 37 E IM54f 9 1 0 83 700 01 83 703 03 83 705 08 83 707 02 83 701 09 83 704 00 83 706 05 83 708 18 83 702 06 83 709 07 83 581 76 83 581 68...

Page 37: ......

Page 38: ...Eschmann Equipment Peter Road Lancing West Sussex BN15 8TJ England Tel 44 0 1903 753322 Fax 44 0 1903 875789 www eschmann co uk...

Page 39: ...for use with active electrode socket 26 black 26 4mm active electrode socket USER 1 footswitch or fingerswitch activation red 27 Fingerswitch sockets for use with active electrode socket 28 black 28...

Reviews: