eschmann TD830 Instructions For Use Manual Download Page 18

P18/37

E-IM54f

pressing a fingerswitch cut or coag button or by pressing
the appropriate footswitch, yellow for cut, blue for coag.
Check that the relevant power activated indicators
illuminate, yellow (9) or blue (22) for cut or coag respectively
and that the audible running tone can be heard.

Notes:

  ♦

The volume of the running tone and the selector
button ‘bleep’ can be adjusted with the volume control
(33) on the back of the unit. This should be adjusted
using a screwdriver of suitable size.

  ♦

Coagulation output is obtained only by pressing the
fingerswitch blue coag button or the blue footswitch.

  ♦

Cut output is obtained only by pressing the fingerswitch
yellow cut button or the yellow footswitch.

5.16 If the bipolar output is to be used ensure that the
bipolar electrode is safe and activate the output by pressing
the white footswitch connected (either electric or
pneumatic). Check that the blue power activated indicator
(7) illuminates and that the audible running tone can be
heard.

5.17 Check that the following alarms are functioning
correctly by the test methods that follow:

PAM - To test the PAM alarm attach the divided plate

electrode to the divided plate cable with the divided
plate cable plugged into the unit. The PAM alarm
should operate with the plate electrode not attached
to the patient.

PCM - To test the PCM alarm remove the non-divided plate

electrode from the non-divided plate cable with the
non-divided plate cable plugged into the unit. The
PCM alarm should operate.

PEM - To test the PEM alarm with the plate electrode

attached to the unit allow the plate electrode to contact
the equipotentiality point (34) at the rear of the unit
using a bare metal part of the plate and not a part
covered by adhesive. With cut and coag power levels
both adjusted to zero (for safety) the PEM alarm
should operate when a USER 1 footswitch or
fingerswitch (whichever is connected to the unit) or a
USER 2 fingerswitch is pressed .

Note: This test is not possible with a capacitive plate
electrode.

 -To test the multiple activation alarm press both a cut

and coag footswitch or fingerswitch control, the alarm
should operate.

The other alarms (PVM, EPM and !) cannot easily be
checked by the user without special test equipment or by
using procedures that could lead to accidental burns. The
function of the EPM alarm should be checked regularly, in
accordance with the Service Manual (see section 1.8).

OPERATING INSTRUCTIONS

Note: For numbers in brackets refer to Illustration 11 which
folds out at the end of this manual.

  ♦

Unit settings (e.g. range selection and mode selection)
are stored in the unit's memory during power-off
periods and will not need reselecting when power is
reapplied. Power settings will also be maintained
during power ‘off’ so long as the control knobs are not
moved.

  ♦

The audible running tones operate at different
frequencies for each of the three output groups (bipolar,
monopolar cut and monopolar coag) as detailed in the
‘Technical data’ section.

If a monopolar user activates power at the same time
(but after) a bipolar user the tone will continue at the
bipolar frequency until the bipolar user stops activation
when the monopolar user’s frequency will start
(assuming the monopolar user is still activating the
unit). If the monopolar user only activates the unit for
a short period during a long activation by the bipolar
user then the monopolar user’s frequency will not be
heard. The same principles apply if a monopolar user
is activating power and a bipolar user starts to activate
power at the same time.

Monopolar standby mode

5.18 If the unit is in monopolar standby mode only bipolar
operation is available, monopolar is disabled and the
monopolar digital displays (8 and 23) will not be seen. The
unit is switched in and out of standby mode be pressing
the monopolar standby button (32). The switch ‘toggles’
standby mode ‘on’ or ‘off’ each time the button is pressed,
monopolar standby mode is selected when the digital
displays for monopolar (8 and 23) are not illuminated.

Monopolar mode

USER 1 and/or USER 2 selection

5.19 The TD830 can be used by two monopolar operators
USER 1 and USER 2 independently of each other but not
simultaneously. Each user can have their own active output
under their own control operated by footswitch or
fingerswitch for USER 1 and fingerswitch only for USER 2
(see sections  5.3 to 5.7)

5.20 USER 1 and USER 2  can be using either cut or
coag but the mode set within cut or coag must be the same
and the power level set must similarly be the same (unless
changed between each USER’s application of power).

5.21 Because USER 1 and USER 2 cannot both be
delivering power at the same time, power is available on a
‘first-come-first-served’ basis. (i.e. if USER 1 is using power
and USER 2 presses a fingerswitch to demand power, none
will be available and visa versa). Also note that power will

Summary of Contents for TD830

Page 1: ...TD830 ELECTROSURGICAL UNIT Instructions for use E IM54f 698257...

Page 2: ...Sussex BN15 8TJ England Tel 44 0 1903 765040 Fax 44 0 1903 875711 Overseas Customers Contact your local distributor In case of doubt contact Eschmann Equipment Patents and Trade marks The ESCHMANN nam...

Page 3: ...ns 14 Electrosurgical interference 14 Using two electrosurgical units 15 Using a flexible endoscope 15 Servicing 15 References 15 5 OPERATION Routine check 16 Connection of accessories 16 Monopolar 16...

Page 4: ...is most important that fuses of the correct type size and rating are installed see Technical Data CAUTION Read these Instructions for use carefully before using this electrosurgical unit and note ALL...

Page 5: ...25V min T10A Fuse type 20 mm Bipolar Carrier frequency 785kHz nominal square wave Power control Variable pulse group modu lation set by front panel control Amplitude set by micro macro range buttons L...

Page 6: ...rd edition General classification Class 1 Type CF Defibrillator proof Drip proof IPX 1 Patient leakage risk current always less than 100 microamps to earth ground from all patient circuits at 230V 50H...

Page 7: ...he symbol on the rear panel serial plate warns the user to read the accompanying documents the Instructions for use The symbol above sockets 25 26 27 and 28 denotes connection socket for a two button...

Page 8: ...Plate voltage monitor alarm has been activated EPM Indicates the Excess power monitor alarm has been activated Indicates the Multiple activation alarm has been activated Indicates an internal fault ha...

Page 9: ...heplateelectrodeisreliably attached to the patient s body and as close to the operating field as possible Store temporarily unused active electrodes such that they are isolated from the patient e g us...

Page 10: ...f the delivered power exceeds the allowed value by more than a certain margin Patient Earth Monitor PEM 3 5 The PEM circuit detects a low impedance path between the patient and earth It also reduces t...

Page 11: ...f the bipolar forceps and the tissue held between them is directly in view of the surgeon at all times 4 7 A plate electrode is not required and safety is increased even further as two power range out...

Page 12: ...ined throughout the surgical procedure The following points should also be noted Do not apply any plate electrode over bony protuberances metal implants folds of skin scar tissue hairy areas any form...

Page 13: ...ited number of re sterilization cycles Regular testing and inspection is therefore essential Do not sterilize and reuse single use types 4 32 Do not connect footswitches to the USER 1 output of the el...

Page 14: ...te electrode are kept a few inches away from the pacemaker pulse generator and its electrode leads while to avoid burns to the myocardium at the electrode site or possible ventricular fibrillation the...

Page 15: ...the following notes should be considered when writing a procedure for the use of this electrosurgical unit with an endoscope a To avoid the risk of potential burns caused by h f voltages on the body o...

Page 16: ...a supply of new plate electrodes is readily available Plate electrode With reusable types examine the platecableandplugforsignsofwear Ensurethatthe insulationoftheconnectingcableisingoodcondition Acce...

Page 17: ...olar electric socket 37 or a white pneumatic footswitch REF 83 581 68 Fig 10 to the bipolar pneumatic socket 36 Do not connect both types of footswitch at the same time 5 9 If the electrosurgical unit...

Page 18: ...uld be checked regularly in accordance with the Service Manual see section 1 8 OPERATING INSTRUCTIONS Note For numbers in brackets refer to Illustration 11 which folds out at the end of this manual Un...

Page 19: ...obtained with the normal power setting check the following before increasing power The plate electrode is correctly positioned and is in contact with the patient Theplateelectrodecableisintact andispl...

Page 20: ...non contact coagulation of tissue that tends to adhere to active electrodes Note Spray coag indicated with the bright green LED is not recommended for laparoscopic and endoscopic procedures where its...

Page 21: ...de or press macro bipolar selector touch button 5 to select macro bipolar mode The green indicators 2 or 4 will illuminate to show which bipolar mode has been selected 5 49 Set power output control 1...

Page 22: ...nit is in standby mode only the bipolar display will flash 5 57 If more than one output demand is received at the same time such as two footswitches or a footswitch and a fingerswitch button being pre...

Page 23: ...s been activated Remedy i If unit PVM alarm activates check for accidental direct contact between the active electrode and an earthed object ii Check that the plate electrode is correctly attached to...

Page 24: ...ation 40 to 60 TCRE TCAE Blend Normal endometrium 55 to 70 Transcervical Resection of Specialist cut Spongio endometrium 40 to 60 the Endometrium Ablation Blend Suspect fibroid inv 65 to 80 Spray coag...

Page 25: ...Power output watts Load resistance 200 ohms MONOPOLAR CUT POWER v CONTROL 0 50 100 150 200 250 300 0 25 50 75 100 125 150 175 200 225 250 Output control setting typical pow er Power output watts Load...

Page 26: ...0 ohms MONOPOLAR BLEND POWER v CONTROL 0 40 80 120 160 200 0 20 40 60 80 100 120 140 160 180 200 Output control setting typical power Power output watts Load resistance 400 ohms MONOPOLAR BLEND DIAGRA...

Page 27: ...ance 200 ohms MONOPOLAR SPECIALIST CUT POWER v CONTROL 0 50 100 150 200 250 300 350 0 30 60 90 120 150 180 210 240 270 300 Output control setting typical pow er Power output watts Load resistance 200...

Page 28: ...MONOPOLAR PINPOINT COAG POWER v CONTROL 0 20 40 60 80 100 120 0 10 20 30 40 50 60 70 80 90 100 Output control setting typical pow er Power output watts Load resistance 400 ohms MONOPOLAR PINPOINT COA...

Page 29: ...alue up to 10 Peak output voltage volts Open circuit MONOPOLAR SPRAY COAG VOLTAGE v CONTROL 0 500 1000 1500 2000 2500 3000 3500 4000 0 7 5 15 22 5 30 37 5 45 52 5 60 67 5 75 Output control setting typ...

Page 30: ...tput voltage volts Open circuit BIPOLAR MACRO POWER v CONTROL 0 10 20 30 40 50 0 1 2 3 4 5 6 7 8 9 10 Output control setting numerical value up to 10 Power output watts Load resistance 100 ohms BIPOLA...

Page 31: ...pow er Peak output voltage volts Open circuit BIPOLAR MICRO POWER v CONTROL 0 4 8 12 16 0 1 5 3 4 5 6 7 5 9 10 5 12 13 5 15 Output control setting typical pow er Power output watts Load resistance 10...

Page 32: ...Eschmann trained personnel otherwise the warranty for the equipment could be infringed 6 5 If the unit is to be stored for any length of time it should be plugged into the mains and switched on to ch...

Page 33: ...sposable 8 83 588 85 4 2mm pack of 5 8 REF Description Illustration No Colposcopy electrodes LLETZ cone biopsy 83 583 51 large loop 12mm pack of 5 8 83 584 19 large loop 15mm pack of 5 8 83 584 27 lar...

Page 34: ...34 37 E IM54f 83 122 42 83 122 46 83 122 49 83 207 56 83 122 93 83 109 39 83 109 47 83 581 09 83 581 13 83 581 17 83 581 25 83 140 20 4mm 8mm 83 141 01 83 141 36 83 141 28 83 141 44 1 2 3 4 5 83 109 7...

Page 35: ...8 m m 6 7 8 83 582 14 83 582 27 83 582 30 83 582 49 83 588 69 83 585 04 Disposable 83 585 11 Disposable 83 582 22 83 583 11 83 585 23 Disposable 83 588 77 83 588 80 83 583 51 83 584 19 83 584 27 83 5...

Page 36: ...P36 37 E IM54f 9 1 0 83 700 01 83 703 03 83 705 08 83 707 02 83 701 09 83 704 00 83 706 05 83 708 18 83 702 06 83 709 07 83 581 76 83 581 68...

Page 37: ......

Page 38: ...Eschmann Equipment Peter Road Lancing West Sussex BN15 8TJ England Tel 44 0 1903 753322 Fax 44 0 1903 875789 www eschmann co uk...

Page 39: ...for use with active electrode socket 26 black 26 4mm active electrode socket USER 1 footswitch or fingerswitch activation red 27 Fingerswitch sockets for use with active electrode socket 28 black 28...

Reviews: