Image Optimization
Command
Function
Action
D-STEER
Doppler orientation
Adjust the CFM box
orientation to optimize cos
Θ
Doppler
AUTOADJUST
Automatic adjustment
Activate it
SCALE
Pulse Repetition Frequency
Decrease the PRF as much as
possible
FREQUENCY
Probe transmission frequency
Increase the frequency as
much as possible
BASELINE
Baseline
Move the baseline to
eliminate the aliasing
FILTER
Wall filter
Reduce the filter as much as
possible
Command
Function
Action
D-STEER
Doppler orientation
Adjust the CFM box
orientation to optimize cos
Θ
Doppler
AUTOADJUST
Automatic adjustment
Activate it
SCALE
Pulse Repetition Frequency
Decrease the PRF as much as
possible
FREQUENCY
Probe transmission frequency
Increase the frequency as
much as possible
BASELINE
Baseline
Move the baseline to
eliminate the aliasing
FILTER
Wall filter
Reduce the filter as much as
possible
Tips to
Optimize Slow Flows
Tips to
Optimize Deep Flows