Tips
in
B-Mode
Tips to
Optimize Superficial Imaging
Command
Function
Action
FUNDAMENT/TEI
Probe transmission frequency
Select
RES
option
ZOOM
Magnification of the region
of interest
Enlarge the area as much as
possible
FOCUSES #
Number of transmission
focuses
Increase the focuses
FOCUS POS
Focus position
Move the focuses to the
desired area
MVIEW
(with Linear and
Convex probes)
Acquisition of several
bidimensional images with
different steering angles
Increase to visualize organ
borders and structured
margins (curved or irregular
borders)
Command
Function
Action
FUNDAMENT/TEI
Probe transmission frequency
Select
PEN
option
FOCUS POS
Focus position
Move the focus to the
desired area
TGC
Potentiometers
Amplification of individual
areas of the sector
Move to right the lower
TGC cursor
IMAGING GAIN
Amplification of overall
received echo signals
(General gain)
Turn clockwise for visualize
deep structures
POWER
Transmission power
Increase the power
Tips to
Optimize Deep Imaging